Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens Cios Alpha VA20-mobile X-Ray system Material # 10308191 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.

Recall number
Z-1957-2020
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens Cios Alpha VA20-mobile X-Ray system Material # 10308191 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.
Status
Terminated
Initiated
2020-04-20
Posted
2020-05-04
Terminated
2021-05-10
Reason
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector
Root cause
Process design
Action
Siemens Healthcare issued A Customer Safety Advisory Notice dated 4/20/20 via AX030/20/S. Letter states reason for recall, health risk and action to take: We strongly recommend that you do not disconnect the main unit from the monitor trolley when the system displays the system error message Err 16305 / 80: Confirm this error and repeat your last action until the problem has been corrected by a service engineer. Contact our service organization at 1-800-888-7436. Siemens is currently developing a solution to eliminate the root cause of this problem.
Quantity
4 units
Distribution
US Nationwide distributions.
Lot, serial or model codes
Serial Numbers: 12425, 109, 10063, 12946
510(k) numbers
["K181560"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register