Roche Molecular Systems, Inc.: MagNA Pure 24 System, GMMI: 07290519001 Product Usage: The MagNA Pure 24 System is a fully automated clinical sample extractor. It extracts nucleic acids from various sample types, sample tubes, and volume inputs for multiple downstream applications. The main features of the system are : Scalable purification of 1 to 24 samples Primary/secondary tube handling Inventory and sample management via barcodes Efficient order management via host The MagNA Pure 24 System offers 4 run types: Purification (rack) run Purification (cartridge) run Post Elution run Sample transfer run

Recall number
Z-1956-2018
Recalling firm
Roche Molecular Systems, Inc.
Product
MagNA Pure 24 System, GMMI: 07290519001 Product Usage: The MagNA Pure 24 System is a fully automated clinical sample extractor. It extracts nucleic acids from various sample types, sample tubes, and volume inputs for multiple downstream applications. The main features of the system are : Scalable purification of 1 to 24 samples Primary/secondary tube handling Inventory and sample management via barcodes Efficient order management via host The MagNA Pure 24 System offers 4 run types: Purification (rack) run Purification (cartridge) run Post Elution run Sample transfer run
Status
Terminated
Initiated
2018-05-02
Posted
not on file
Terminated
2019-04-30
Reason
Cross-contamination of samples has been reported when running the existing protocol on the system with the rack-based purification run.
Root cause
Software in the Use Environment
Action
Customers were notified of the field action by letter on 4/27/18. The letter included the following required actions: " If you used the MagNA Pure 24 Pathogen 200 and/or Pathogen 1000 protocols to extract nucleic acids for an assay for a chronic infectious disease (e.g., Hepatitis C Virus), then suspected false positives should be evaluated further following local procedures and retested as necessary. " Complete the attached fax form and fax it to 1-844-861-6232 or email it to roche6612@stericycle.com. " File this Urgent Medical Device Correction (UMDC) for future reference.
Quantity
163 units
Distribution
Worldwide Distribution - US Nationwide in the states of: CA, GA, HI, IN, MD, ME, MI, MN, NC, NY, PA, VI, and WI. The products were distributed to the following foreign countries: Austria, Belgium, Brazil, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, France, Germany, Guatemala, Hong Kong, Hungary, India, Israel, Italy, Japan, Mexico, Netherlands, Norway, Poland, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, UAE, United Kingdom, Vietnam.
Lot, serial or model codes
Device Identifier: 07613336106174
510(k) numbers
not on file
PMA numbers
not on file
Product code
JJH
Firm FEI
2243471
Firm city
S Branchburg
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source