Roche Molecular Systems, Inc.: MagNA Pure 24 System, GMMI: 07290519001 Product Usage: The MagNA Pure 24 System is a fully automated clinical sample extractor. It extracts nucleic acids from various sample types, sample tubes, and volume inputs for multiple downstream applications. The main features of the system are : Scalable purification of 1 to 24 samples Primary/secondary tube handling Inventory and sample management via barcodes Efficient order management via host The MagNA Pure 24 System offers 4 run types: Purification (rack) run Purification (cartridge) run Post Elution run Sample transfer run
- Recall number
- Z-1956-2018
- Recalling firm
- Roche Molecular Systems, Inc.
- Product
- MagNA Pure 24 System, GMMI: 07290519001 Product Usage: The MagNA Pure 24 System is a fully automated clinical sample extractor. It extracts nucleic acids from various sample types, sample tubes, and volume inputs for multiple downstream applications. The main features of the system are : Scalable purification of 1 to 24 samples Primary/secondary tube handling Inventory and sample management via barcodes Efficient order management via host The MagNA Pure 24 System offers 4 run types: Purification (rack) run Purification (cartridge) run Post Elution run Sample transfer run
- Status
- Terminated
- Initiated
- 2018-05-02
- Posted
- not on file
- Terminated
- 2019-04-30
- Reason
- Cross-contamination of samples has been reported when running the existing protocol on the system with the rack-based purification run.
- Root cause
- Software in the Use Environment
- Action
- Customers were notified of the field action by letter on 4/27/18. The letter included the following required actions: " If you used the MagNA Pure 24 Pathogen 200 and/or Pathogen 1000 protocols to extract nucleic acids for an assay for a chronic infectious disease (e.g., Hepatitis C Virus), then suspected false positives should be evaluated further following local procedures and retested as necessary. " Complete the attached fax form and fax it to 1-844-861-6232 or email it to roche6612@stericycle.com. " File this Urgent Medical Device Correction (UMDC) for future reference.
- Quantity
- 163 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of: CA, GA, HI, IN, MD, ME, MI, MN, NC, NY, PA, VI, and WI. The products were distributed to the following foreign countries: Austria, Belgium, Brazil, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, France, Germany, Guatemala, Hong Kong, Hungary, India, Israel, Italy, Japan, Mexico, Netherlands, Norway, Poland, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, UAE, United Kingdom, Vietnam.
- Lot, serial or model codes
- Device Identifier: 07613336106174
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JJH
- Firm FEI
- 2243471
- Firm city
- S Branchburg
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28