Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Navigation, Inc.: StealthStation S7 Framelink Software version 5.4. Aid for locating anatomical structures and planning surgical trajectories in open and percutaneous procedures.

Recall number
Z-1956-2013
Recalling firm
Medtronic Navigation, Inc.
Product
StealthStation S7 Framelink Software version 5.4. Aid for locating anatomical structures and planning surgical trajectories in open and percutaneous procedures.
Status
Terminated
Initiated
2013-07-15
Posted
2013-08-13
Terminated
2014-01-29
Reason
Medtronic Navigation, Inc. has initiated a field correction to their Framelink software version 5.4 which can be used on Medtronic Navigation StealthStation 57 and i7 systems or a Medtronic Planning Station due to the potential of discrepancy between the exam image and the displayed measurement or overlay under certain conditions of use.
Root cause
Software design
Action
A Field Safety Notice was sent on 7/15/13 to users informing them that the Framelink software version 5.4 has been revised to correct the software anomalies described in this letter. A new version will be installed on their system(s) by a Medtronic Navigation representative. Until the software upgrade has been installed on their system, do not use high resolution examinations (greater than 256 x 256). If users are not sure of the exam resolution being used, firm recommend contacting the hospital imaging department or local Medtronic Navigation representative for assistance. Medtronic representatives will visit all identified hospitals and upgrade all systems to version 5.4.1. This will account for all consignees.
Quantity
572
Distribution
Worldwide Distribution.
Lot, serial or model codes
Lot number(s): Not applicable; All kits containing the Framelink Software version 5.4 are affected by this corrective action.
510(k) numbers
["K992927"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000517638
Firm city
Louisville
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register