FDA Medical Device Recalls (openFDA)
Dynarex Corporation: Dynarex CGA870 All Brass Oxygen Regulator, All Brass - 0-25 LPM, Barb & 2-DISS Outlet, UPC: 616784522715, Model Number 5227 A pressure regulator is a device, often called a pressure-reducing valve, that is intended for medical purposes and that is used to convert a medical gas pressure from a high variable pressure to a lower, more constant working pressure. This device includes mechanical oxygen regulators.
- Recall number
- Z-1955-2018
- Recalling firm
- Dynarex Corporation
- Product
- Dynarex CGA870 All Brass Oxygen Regulator, All Brass - 0-25 LPM, Barb & 2-DISS Outlet, UPC: 616784522715, Model Number 5227 A pressure regulator is a device, often called a pressure-reducing valve, that is intended for medical purposes and that is used to convert a medical gas pressure from a high variable pressure to a lower, more constant working pressure. This device includes mechanical oxygen regulators.
- Status
- Terminated
- Initiated
- 2018-05-08
- Posted
- not on file
- Terminated
- 2018-09-18
- Reason
- The device is equipped with a dial that has an inaccurate flow rate indicator.
- Root cause
- Process control
- Action
- The Firm sent an Urgent Medical Device Recall letter on May 8, 2018. Customers were instructed : Customer/User should stop distributing/using product immediately. Customer/User is to complete the attached Medical Device Recall Return Response Acknowledgement and Receipt Form and return immediately using the fax number or email address on the form. After returning the form, please contact Dynarex Returned Goods Representative (see contact information on attached for or under (8) below) to return the product and to receive credit for the product. For further questions, please call ( 845 ) 365-8200.
- Quantity
- 15 devices
- Distribution
- US Distribution to the states of : MA and NJ
- Lot, serial or model codes
- Lot number: 37979
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- CAN
- Firm FEI
- 2431014
- Firm city
- Orangeburg
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28