Skip to the content

FDA Medical Device Recalls (openFDA)

Draeger Medical Systems, Inc.: Stabilet Infant Warming System, model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200.

Recall number
Z-1955-2009
Recalling firm
Draeger Medical Systems, Inc.
Product
Stabilet Infant Warming System, model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200.
Status
Terminated
Initiated
2009-07-17
Posted
2009-08-25
Terminated
2011-06-01
Reason
Hot particles fell into the infant bassinet beneath a Radiant Infant Warmer. The resulting fire injured an infant. These devices are obsolete.
Root cause
Equipment maintenance
Action
On 7/20/09, Draeger Medical, the recalling firm, issued a press release to medical media outreach organizations and an Urgent Medical Device letter to all known customers. This information explained the incident and the need to remove and replace the units in question as soon as feasibly possible.
Quantity
6489 units
Distribution
Worldwide distribution: USA, Brazil, Canada, China, Dominican Republic, France, Korea, Lebanon, Mexico, Nigeria, Panama, Poland, Saudi Arabia, Taiwan, Turkey, United Kingdom, Uruguay, and Venezuela.
Lot, serial or model codes
All units of model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200.
510(k) numbers
["K042607","K850069","K913945"]
PMA numbers
not on file
Product code
FMT
Firm FEI
2510954
Firm city
Telford
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register