FDA Medical Device Recalls (openFDA)
Medrad Inc: Stellant CT Dual Syringe Kits, Catalog Number SDS-CTP-QFT and SDS-CTP-SPK
- Recall number
- Z-1954-2010
- Recalling firm
- Medrad Inc
- Product
- Stellant CT Dual Syringe Kits, Catalog Number SDS-CTP-QFT and SDS-CTP-SPK
- Status
- Terminated
- Initiated
- 2010-05-24
- Posted
- 2010-07-09
- Terminated
- 2010-10-13
- Reason
- Kits contain a T connector that may be susceptible to cracking or breaking or leaking. This could cause product leakage or, less likely, patient injury.
- Root cause
- Other
- Action
- The recalling firm issued an Urgent Medical Device Field Correction on 5/24/10 requesting customers to check their inventory for the affected lots and complete the fax back form with the amount of product in stock. The manufacturer plans to make arrangements to provide individually packaged low-pressure connector tubing to use with the product in the event that cracking or breaking and leaking of the T connector is observed. This requires vigilance by the customers over the life of all of the devices. For questions concerning this recall, please contact the company 1-800-633-7231, select option 2, and indicate that you are calling about the Stellant CT Dual Syringe Kit field correction.
- Quantity
- 1,382,889 kits
- Distribution
- nationwide and Worldwide distribution includes AU, BE, BG, BM, BR, CA, CH, CI, CL, CN, CO, CS, DE, DK, EC, EG, FI, FR, GB, GT, HN, IN, IT, JM, JP, KW, KY, MU, MX, MY, NL, OM, PA, PH, PK, PR, SE, SG, TW, YE, ZA.
- Lot, serial or model codes
- 99725, 99726, 99727, 99728, 99729, 99730, 99731, 99733, 99734, 99747, 99748, 99749, 99750, 99753, 99754, 99755, 99757, 99766, 99767, 99768, 99772, 99775, 99776, 99777, 99778, 99813, 100053, 100464, 100465, 100466, 100467, 100471, 100472, 100473, 100474, 100475, 100476, 100477, 100478, 100479, 100488, 100512, 100513, 100514, 100515, 100516, 100517, 100518, 100519, 100520, 100521, 100523, 100524, 100525, 100526, 100528, 100529, 100531, 100532, 100533, 100535, 100537, 100538, 100539, 100541, 100542, 100543, 100546, 100547, 100549, 100550, 100551, 100552, 100553, 100555, 100558, 100559, 100560, 100561, 100564, 100567, 100568, 100569, 100570, 100897, 101378, 101379, 101380, 101386, 101387, 101391, 101394 with exp dates between 2/2014 to 4/2014.
- 510(k) numbers
- ["K023183"]
- PMA numbers
- not on file
- Product code
- DXT
- Firm FEI
- 3006316319
- Firm city
- Warrendale
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28