Skip to the content

FDA Medical Device Recalls (openFDA)

Gyrus Acmi, Incorporated: Gyrus ACMI Reusable Active Cord for Bovie/ValleyLab Generators, Catalog No C650-129A (00821925008700) The cord consists of a single conductor stranded wire terminated with plugs that connect to a generator on one end and a plug that connects to a Gyrus ACMI working element on the other end. They are provided non-sterile and are reusable. The cord is designed to be used with Gyrus ACMI GOCUS 640, Bovie, and other general purpose monopolar electrosurgical generators and appropriate electrodes and endoscopes.

Recall number
Z-1952-2018
Recalling firm
Gyrus Acmi, Incorporated
Product
Gyrus ACMI Reusable Active Cord for Bovie/ValleyLab Generators, Catalog No C650-129A (00821925008700) The cord consists of a single conductor stranded wire terminated with plugs that connect to a generator on one end and a plug that connects to a Gyrus ACMI working element on the other end. They are provided non-sterile and are reusable. The cord is designed to be used with Gyrus ACMI GOCUS 640, Bovie, and other general purpose monopolar electrosurgical generators and appropriate electrodes and endoscopes.
Status
Terminated
Initiated
2018-04-17
Posted
not on file
Terminated
2023-09-18
Reason
There is a potential for the cord to spark and cause a fire.
Root cause
Device Design
Action
Notification letters were distributed on April 17, 2018. The letters instruct customers to perform the following: Immediately assess any affected product you have in stock and replace the IFU with the revised IFU. Also ensure that any potential users of the devices are aware of the additional warning in the revised IFU. Future versions of the IFU will include the warning. Remove and discard all cords in use greater than one year. If you need additional copies of the IFU, call your Olympus customer service representative at 1-888-524-7266 to arrange for them to be provided. Please note on the enclosed Reply Form that you have received this information. Fax the completed Reply Form to 484-896-7128 regardless of whether you have any affected inventory at your facility. In addition, if you may have further distributed this product, please identify your customers, inform them at once of this action, provide the revised IFU and appropriately document your notification process. Your notification to your customers may be enhanced by including a copy of this letter. For further questions, please call (508) 804-2600.
Quantity
6447
Distribution
Worldwide Distribution - USA (nationwide) Distribution to the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, and WY., and to the countries of : Japan, Canada, Mexico, Australia, and EU.
Lot, serial or model codes
All lots manufactured before December 7, 2017  UDI (00821925008700)
510(k) numbers
["K890328"]
PMA numbers
not on file
Product code
FAS
Firm FEI
3003790304
Firm city
Southborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register