FDA Medical Device Recalls (openFDA)
Helena Laboratories, Corp.: ColoCARE, Cat. No. 5651, packaged 50 envelopes/box, For In-Vitro Diagnostic Use. The firm name on the label is Helena Laboratories, Beaumont, TX.
- Recall number
- Z-1950-2019
- Recalling firm
- Helena Laboratories, Corp.
- Product
- ColoCARE, Cat. No. 5651, packaged 50 envelopes/box, For In-Vitro Diagnostic Use. The firm name on the label is Helena Laboratories, Beaumont, TX.
- Status
- Open, Classified
- Initiated
- 2018-08-08
- Posted
- not on file
- Terminated
- not on file
- Reason
- The positive control on some tests of the lots did not appropriately react within the specified timeframe to verify the test was functioning correctly.
- Root cause
- Nonconforming Material/Component
- Action
- The recalling firm made notifications beginning 8/8/2018 via phone and letter delivered via email. Customers were to dispose of the product.
- Quantity
- 328/50-pouch kits
- Distribution
- Distribution was made to AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MO, MS, MT, NC, ND, NE, NV, NY, OH, OK, PA, SD, TN, TX, VA, WA, and WY. There was no foreign/military/government distribution.
- Lot, serial or model codes
- Kit lots 2-17-5651, exp. 8/31/2018, and 3-17-5651, exp. 9/30/2018. These kit lots contain ColoCARE pad lot numbers 1-17-551039, exp. 8/31/2018, and 2-17-551039, exp. 9/30/2018.
- 510(k) numbers
- ["K882799"]
- PMA numbers
- not on file
- Product code
- KHE
- Firm FEI
- 1618982
- Firm city
- Beaumont
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28