Helena Laboratories, Corp.: ColoCARE, Cat. No. 5651, packaged 50 envelopes/box, For In-Vitro Diagnostic Use. The firm name on the label is Helena Laboratories, Beaumont, TX.

Recall number
Z-1950-2019
Recalling firm
Helena Laboratories, Corp.
Product
ColoCARE, Cat. No. 5651, packaged 50 envelopes/box, For In-Vitro Diagnostic Use. The firm name on the label is Helena Laboratories, Beaumont, TX.
Status
Open, Classified
Initiated
2018-08-08
Posted
not on file
Terminated
not on file
Reason
The positive control on some tests of the lots did not appropriately react within the specified timeframe to verify the test was functioning correctly.
Root cause
Nonconforming Material/Component
Action
The recalling firm made notifications beginning 8/8/2018 via phone and letter delivered via email. Customers were to dispose of the product.
Quantity
328/50-pouch kits
Distribution
Distribution was made to AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MO, MS, MT, NC, ND, NE, NV, NY, OH, OK, PA, SD, TN, TX, VA, WA, and WY. There was no foreign/military/government distribution.
Lot, serial or model codes
Kit lots 2-17-5651, exp. 8/31/2018, and 3-17-5651, exp. 9/30/2018. These kit lots contain ColoCARE pad lot numbers 1-17-551039, exp. 8/31/2018, and 2-17-551039, exp. 9/30/2018.
510(k) numbers
["K882799"]
PMA numbers
not on file
Product code
KHE
Firm FEI
1618982
Firm city
Beaumont
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register