Beckman Coulter Inc.: Access Total T4 Reagent Kit, Part Number: 33800, Lot Numbers (s): All reagent pack lots The Access Total T4 assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total thyroxine (T4) levels in human serum and plasma using the Access Immunoassay Systems. Total T4 assay results are used for diagnosis and confirmation of thyroid disorders such as hyper- and hypothyroidism.

Recall number
Z-1949-2011
Recalling firm
Beckman Coulter Inc.
Product
Access Total T4 Reagent Kit, Part Number: 33800, Lot Numbers (s): All reagent pack lots The Access Total T4 assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total thyroxine (T4) levels in human serum and plasma using the Access Immunoassay Systems. Total T4 assay results are used for diagnosis and confirmation of thyroid disorders such as hyper- and hypothyroidism.
Status
Terminated
Initiated
2011-03-10
Posted
2011-04-12
Terminated
2012-08-10
Reason
The recall was initiated because the concentration of total T4 in samples from pregnant women is erroneously low (20%) when measured using the Access Total T4 assay. Erroneously low results which are actually high (outside of the normal reference range) could mask hyperthyroidism. Failure to diagnose or treat hyperthyroid disease associated with pregnancy could result in serious, permanent harm
Root cause
Other
Action
Beckman Coulter sent an Urgent Product Correction letter dated March 20, 2011, to all affected consignees. The letter identified the product, the problem and the action to be taken by the consignee. Consignees were issued the following instructions: (1) If you suspect pregnant patient results were evaluated at your laboratory using only Total T4 to report thyroid status, consider reviewing Total T4 results reported in the last 9 months. Consult with your Laboratory Medical Director and requesting physician to resolve any concerns you may have regarding previously reported patient results during this time period. (2) Do not use the Access Total T4 assay as the only marker for evaluating pregnant patients for thyroid disorders. (3) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded Access Total T4 reagent kits (Part Number 33800) to another laboratory, please provide a copy of this letter to them. (4) Please complete and return the enclosed Response Form within 10 days so we are assured you have received this important communication. For questions regarding this notice, consignees were instructed to contact Beckman Coulter Customer Support Center. (1) Via our website, http://www.beckmancoulter.com/customersupport/support (2) Via phone, call 1-800-854-3633 in the United States and Canada (3) Outside the United States and Canada please contact your local Beckman Coulter
Quantity
151,740 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Argentina Armenia, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Djibouti, Egypt, El Salvador, Ethiopia, France, Germany, Greece, Guatemala, Hong Kong, India, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Kuwait, Lebanon, Libyan Arab Jamahiriya, Liechtenstein, Malaysia, Mexico, Republic of Moldova, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Pakistan, Panama, Philippines, Poland, Puerto Rico, Reunion, Romania, Russian Federation, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Yemen, and Zambia
Lot, serial or model codes
Lot Numbers (s): All reagent pack lots
510(k) numbers
["K023369","K926475"]
PMA numbers
not on file
Product code
KLI
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register