Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics: Colorimetric (CM)/ Rate (RT) incubator wear pads. Product Codes: 1) 6801375, 2) 6801890, 3) 6802445, 4) 6802413 Product Usage: The CM/RT incubator processes Microslide reagents with the following VITROS systems: 1) VITROS 5,1 FS Chemistry System, 2) VITROS 5,1 FS Chemistry System Refurbished, 3) VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), VITROS 5600 Integrated System.

Recall number
Z-1948-2013
Recalling firm
Ortho-Clinical Diagnostics
Product
Colorimetric (CM)/ Rate (RT) incubator wear pads. Product Codes: 1) 6801375, 2) 6801890, 3) 6802445, 4) 6802413 Product Usage: The CM/RT incubator processes Microslide reagents with the following VITROS systems: 1) VITROS 5,1 FS Chemistry System, 2) VITROS 5,1 FS Chemistry System Refurbished, 3) VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), VITROS 5600 Integrated System.
Status
Terminated
Initiated
2013-06-06
Posted
not on file
Terminated
2018-05-17
Reason
Ortho Clinical Diagnostics is recalling CM/RT wear pads which are a component in the CM/RT incubator due to becoming worn before their replacement.
Root cause
Nonconforming Material/Component
Action
Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 6, 2013 via FedEx to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that an OCD trained service personnel will contact them and schedule a replacement of the CM/RT wear pads. OCD requested customers to complete and return the enclosed Confirmation of Receipt form no later than June 17, 2013. Customers are asked to place a copy of the notification by each VITROS System in their facility. Foreign affiliates were notified by email on June 6, 2013 of the issue and instructed to notify their consignees of the issue and actions. If you have any questions or need additional information, please call Customer Technical Services at 1-800-421-3311.
Quantity
not on file
Distribution
Worldwide Distribution - USA Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, including Puerto Rico and the countries of Mexico, Singapore, Venezuela, England, Germany, Italy & Spain.
Lot, serial or model codes
Serial Numbers: 1) J34000102 through J34002323, 2) 34000133, 34000193, 4000212, 34000316, 34000318, 34000337, 34000360, 34000420, 34000451, 34000478, 34000495, 34000497, 34000529, 34000538, 34000552, 34000557,  34000596, 34000601, 34000631, 34000642, 34000805, 34000833, 34000851, 34000893, 34000947, 34001172, 34001218, 34001240, 34001243, 34001248, 34001251, 34001274, 34001297, 34001311, 34001317, 34001345, 34001378, 34001389, 34001390, 34001396, 34001422, 34001472, 34001504, 34001569, 34001572, 34001589, 34001610, 34001612, 34001633, 34001668, 34001809,  34001840, 34001850, 34001897, 3) J46000108 through J46000312, 4) J56000110 through J56001779.
510(k) numbers
["K031924","K063144","K081543"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register