Siemens Healthcare Diagnostics, Inc.: Syva EMIT 2000 Theophylline Product Usage: The Syva Emit 2000 Theophylline Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of theophylline in human serum or plasma.

Recall number
Z-1946-2018
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Syva EMIT 2000 Theophylline Product Usage: The Syva Emit 2000 Theophylline Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of theophylline in human serum or plasma.
Status
Terminated
Initiated
2017-11-27
Posted
not on file
Terminated
2019-04-19
Reason
Lot J1 of the Syva EMIT 2000 Theophylline Assay, when run on a Beckman AU Clinical Chemistry System, may exhibit increased imprecision for commercially available controls and for patient samples.
Root cause
Nonconforming Material/Component
Action
Siemens sent an Urgent Medical Device Recall letter dated November 27, 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of affected product, review inventory to determine replacement needs, complete and return Field Correction Effectiveness Check list. For questions contact Siemens Customer Care Center or your local Siemens technical support representative.
Quantity
not on file
Distribution
US Nationwide Distribution
Lot, serial or model codes
Catalog # 4P019UL /SMN#10445324 kit lot J1 (UDI# 00842768001024J120190414)
510(k) numbers
["K913123"]
PMA numbers
not on file
Product code
KLS
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register