Skip to the content

FDA Medical Device Recalls (openFDA)

Axis-Shield Diagnostics, Ltd.: Axis-Shield Liquid Stable (LS) 2-Part HOMOCYSTEINE REAGENT; 100 test format, 1000 test format, and 200 test format; and Private Label: (Distributed by BECKMAN COULTER, for professional use only, on the BECKMAN COULTER AU platforms (AU400, AU480, AU680 and AU5800)) Product Usage: For in-vitro quantitative determination of total homocysteine in human plasma or serum.

Recall number
Z-1946-2015
Recalling firm
Axis-Shield Diagnostics, Ltd.
Product
Axis-Shield Liquid Stable (LS) 2-Part HOMOCYSTEINE REAGENT; 100 test format, 1000 test format, and 200 test format; and Private Label: (Distributed by BECKMAN COULTER, for professional use only, on the BECKMAN COULTER AU platforms (AU400, AU480, AU680 and AU5800)) Product Usage: For in-vitro quantitative determination of total homocysteine in human plasma or serum.
Status
Terminated
Initiated
2015-03-20
Posted
2015-07-01
Terminated
2015-10-27
Reason
Deterioration in performance of the Liquid Stable (LS) 2-Part HOMOCYSTEINE REAGENT resulting in under recovery.
Root cause
Under Investigation by firm
Action
Axis Shield sent an Urgent Field Safety Notice letter dated March 20, 2015 to their customers via email and written notice sent with confirmed delivery receipt. The letter identified the affected product, problem and actions to be taken. Recommended Action to Customer is as follows: Confirm receipt of FSN to Axis-Shield. Do not use the affected Reagent lots referenced in the safety notice. In the event of holding one of the affected lots, immediately contact Axis-Shield Product Support Department to arrange for replacement material and return any defective product to Axis-Shield. Complete reconciliation form. Complete acknowledgement of the safety notice within 10 working days. Ensure safety notice has been distributed and reviewed by all appropriate personnel within your organization. If further distribution of affected lots to another organization has been made, inform them of this Product Recall and provide a copy of this safety notice to them. Retain a copy of the safety notice for their records. For any issues or concerns contact Axis-Shield via e-mail; MDL-AXD-Product- Support@alere.com.
Quantity
52 kits + 75 kits Beckman Coulter Brand OUS
Distribution
Worldwide Distribution - US Nationwide in the states of: KY, WA, CA, and PA and the country of Germany.
Lot, serial or model codes
Part Number/Lot Number: FHBC100/802899574; FHRWR100/802899249; FHRWR1000/802899248; FHRWR200/802899989; FHRWR200/802899250; FHRWCN1/802900104
510(k) numbers
["K083222"]
PMA numbers
not on file
Product code
LPS
Firm FEI
3002808140
Firm city
Dundee
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register