Skip to the content

FDA Medical Device Recalls (openFDA)

Stanley Security Solutions, Inc.: Tabs Professional Monitor Without Pull Cord or wall bracket, Model 25025, Stanley Senior Technologies, Lincoln, NE.

Recall number
Z-1946-2010
Recalling firm
Stanley Security Solutions, Inc.
Product
Tabs Professional Monitor Without Pull Cord or wall bracket, Model 25025, Stanley Senior Technologies, Lincoln, NE.
Status
Terminated
Initiated
2010-03-26
Posted
2010-07-01
Terminated
2011-09-16
Reason
Product might not indicate low battery alarms or exit alarms when powered by a DC power supply.
Root cause
Component change control
Action
The firm made phone calls beginning 3/26/10 to customers who purchased the 12-volt power supply informing them of the recall and instructing them to determine the number of units that will need to be returned for replacement. Important Safety Notice letters dated 3/25/10 were issued via regular mail on 3/26/10 and also providing the reason for the recall, instructions, and customer service contact information to arrange for return of the units.
Quantity
22 units
Distribution
Nationwide Distribution
Lot, serial or model codes
Serial numbers 236920 thru 236941
510(k) numbers
["K873806"]
PMA numbers
not on file
Product code
KMI
Firm FEI
1929691
Firm city
Lincoln
Firm state
NE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register