FDA Medical Device Recalls (openFDA)
Clinical Innovations, LLC: Brand Name: Kiwi Omni Vacuum Delivery System Product Name: Kiwi Omni Vacuum Delivery System Model/Catalog Number: VAC-6000MT Product Description: The Clinical Innovations Kiwi is a disposable vacuum assisted fetal delivery system. It is a sterile, single-patient-use device designed to provide assistance in childbirth under the following conditions: 1) Term pregnancy, 2) Ruptured amniotic membranes, 3) Engaged head, 4) Complete cervical dilation, and 5) Adequately trained or supervised operator.
- Recall number
- Z-1945-2026
- Recalling firm
- Clinical Innovations, LLC
- Product
- Brand Name: Kiwi Omni Vacuum Delivery System Product Name: Kiwi Omni Vacuum Delivery System Model/Catalog Number: VAC-6000MT Product Description: The Clinical Innovations Kiwi is a disposable vacuum assisted fetal delivery system. It is a sterile, single-patient-use device designed to provide assistance in childbirth under the following conditions: 1) Term pregnancy, 2) Ruptured amniotic membranes, 3) Engaged head, 4) Complete cervical dilation, and 5) Adequately trained or supervised operator.
- Status
- Open, Classified
- Initiated
- 2026-03-09
- Posted
- 2026-04-24
- Terminated
- not on file
- Reason
- Due to complaints of device breakage at the traction force gauge to handle joint.
- Root cause
- Under Investigation by firm
- Action
- On 03/10/2026, the firm emailed an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that there is a potential of failure on the handle breaking at the Traction Force Indicator handle joint on the Kiwi Complete Vacuum Delivery System (VAC-6000MT) when angular force is applied to the handle. Customers are instructed to: 1. Identify and quarantine all unused inventory of the affected product. 2. Discontinue distribution of the affected product. 3. Return all unused, affected product to Clinical Innovations (Laborie), regardless of the labeled expiration date. 4. Complete and return the attached Recall Acknowledgement and Receipt Form within 15 business days, even if no inventory remains. Return the forms via email to recalls@Laborie.com. 5. If customers have product to return, a company representative will contact them once Clinical Innovations has receive their Recall Acknowledgement and Receipt Form to issue them an RMA per company policy and arrange for a product return. For questions, call Clinical Innovations (Laborie), at 1-(888)- 268-6222 M-F 8:00 AM- 5:00 PM MT or + (33) 383 22 20 76 M-F 8:00 AM - 5 PM GMT +2 or their Clinical Innovation's (Laborie) service representative.
- Quantity
- 49,175 devices
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AR, AZ, CA, CO, CT, DC, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, MI, MI, MN, MS, NC, NE, NH, NV, NY, OH, PR, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Canada, China, Japan, Czech Republic, and Singapore.
- Lot, serial or model codes
- Model/Catalog Number: VAC-6000MT UDI: 00814247020642 Lot Code: 251327 251333 251328 251330 251370 251388 251545 251546 251582 251583 251680 251681
- 510(k) numbers
- ["K981260"]
- PMA numbers
- not on file
- Product code
- HDB
- Firm FEI
- 1000220650
- Firm city
- Murray
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28