Skip to the content

FDA Medical Device Recalls (openFDA)

Bio-Detek, Inc.: ZOLL Pro-padz Liquid Gel Radiolucent, Part numbers: 8900-2105-01 (12/case) and 8900-2106-01 (single)

Recall number
Z-1945-2019
Recalling firm
Bio-Detek, Inc.
Product
ZOLL Pro-padz Liquid Gel Radiolucent, Part numbers: 8900-2105-01 (12/case) and 8900-2106-01 (single)
Status
Terminated
Initiated
2019-05-28
Posted
not on file
Terminated
2020-06-01
Reason
Some of the electrodes may have been assembled incorrectly, preventing delivery of defibrillation or pacing therapy.
Root cause
Process control
Action
The firm notified customers of the recall on May 28, 2019, via Urgent Device Correction letter. Customers were advised to take the following actions: - Inform users of the problem. -Remove all stock of part numbers 8900-2105-01 and 8900-2106-01 with lot numbers 1719, 1719A , and 1819. - Complete the response form and return to ZOLL. A ZOLL representative will contact you to coordinate the return and replacement of your electrodes. Customers may direct their questions to the firm's 24/7 technical support numbers 1 (800) 348-9011 or +1(978)421-9460
Quantity
1789
Distribution
Distributed to accounts in AZ, CA, CO, CT, FL, LA, MN, NV, OH, PA, TN, UT, and WY. International distribution to Canada, Germany, France, and Italy.
Lot, serial or model codes
Lots 1719, 1719A, and 1819
510(k) numbers
["K960676"]
PMA numbers
not on file
Product code
DTE
Firm FEI
1218058
Firm city
Pawtucket
Firm state
RI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register