FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Siemens AXIOM Artis zeego x-ray, angiographic system
- Recall number
- Z-1945-2013
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Siemens AXIOM Artis zeego x-ray, angiographic system
- Status
- Terminated
- Initiated
- 2013-05-13
- Posted
- 2013-08-09
- Terminated
- 2014-11-06
- Reason
- Siemens issued a customer safety notice about the overlay of system information after a functional check of the brakes on the Artis Zeego system.
- Root cause
- Software design
- Action
- The firm, Siemens, sent an "Important safety notice" by letter on May 13, 2013 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to immediately notify the service department to clarify what further measures need to be taken, if after carrying out the regular brake test, the system can only be moved at a reduced speed. If this device has been sold and is therefore no longer in your possession, send this safety notice to the new owner and provide the new owner to Siemens. Update instruction AX012/13/S was issued, and a software update will be released as AX006/13/S and AX007/13/S which will modify the system software so that it will not be possible for information to be overlaid. If you have any questions call +49 (9191) 18 0.
- Quantity
- 132
- Distribution
- Nationwide distribution: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, LA, MA, MD, MN, MO,MS, NC, NE, NH, NJ, NY, OH, PA, PR, SD, TN, TX, UT, VA, WA, and WI.
- Lot, serial or model codes
- Model number 10280959
- 510(k) numbers
- ["K090745"]
- PMA numbers
- not on file
- Product code
- IZI
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28