Skip to the content

FDA Medical Device Recalls (openFDA)

Impac Medical Systems Inc: MOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An image enabled electronic medical record system used for oncology workflow management. Manufactured by: Impac Medical Systems 100 Mathilda Place, 5th floor Sunnyvale, Ca 94086

Recall number
Z-1944-2012
Recalling firm
Impac Medical Systems Inc
Product
MOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An image enabled electronic medical record system used for oncology workflow management. Manufactured by: Impac Medical Systems 100 Mathilda Place, 5th floor Sunnyvale, Ca 94086
Status
Terminated
Initiated
2012-06-08
Posted
2012-07-06
Terminated
2013-06-19
Reason
Impac Medical became aware of a problem with the recording of the Wedge Motor Units in MOSAIQ when it resumes after a machine terminates/faults, the remaining Wedge MU was being erroneously set to zero.
Root cause
Software design
Action
A User Notice began being distributed on June 8, 2012. This will be followed by contact at each site by IMPAC support to determine if the site is affected by this issue, and if so, the upgraded version will be suggested. The Notice identified the affected device, the problem, the clinical impact, along with the interim mitigations and the final resolution. Users are encouraged to monitor daily delivery and recordings of treatments to be sure that the Wedge Motor Units treatments are properly delivered. The letter also recommends that treatment charts be checked weekly and that users not override any parameters in MOSAIQ until the machine setup has been completed. Users are to contact Elekta Support for the upgrade.
Quantity
11 affected sites
Distribution
Worldwide Distribution -- USA, including the states of VA, CT, GA, NJ and NY and the countries of Belgium, Denmark, France, and the United Kingdom.
Lot, serial or model codes
Version 2.10 - all builds and version 2.20 - builds up to and including 2.20.07A7.
510(k) numbers
["K120067"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1000123805
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register