Skip to the content

FDA Medical Device Recalls (openFDA)

US Endoscopy Group Inc: Velocity" Biopsy Valve, 25 units per box, Product Usage: The disposable Velocity biopsy valve is used to cover the opening of the biopsy inlet port of a flexible gastrointestinal endoscope. The Velocity biopsy valve provides access for endoscopic device passage and exchange, helps maintain insufflation, and minimizes leakage of biomaterial from the accessory port throughout the gastrointestinal endoscopic procedure. The Velocity biopsy valve provides a connection for irrigation through the Velocity irrigation pump.

Recall number
Z-1943-2015
Recalling firm
US Endoscopy Group Inc
Product
Velocity" Biopsy Valve, 25 units per box, Product Usage: The disposable Velocity biopsy valve is used to cover the opening of the biopsy inlet port of a flexible gastrointestinal endoscope. The Velocity biopsy valve provides access for endoscopic device passage and exchange, helps maintain insufflation, and minimizes leakage of biomaterial from the accessory port throughout the gastrointestinal endoscopic procedure. The Velocity biopsy valve provides a connection for irrigation through the Velocity irrigation pump.
Status
Terminated
Initiated
2015-02-17
Posted
2015-07-01
Terminated
2015-09-21
Reason
The company has received reports of structural non conformance at the connection between the valve body and the irrigation tube, resulting in separation of the components during use.
Root cause
Under Investigation by firm
Action
US Endoscopy sent an Urgent Recall Notification Letter dated to their customers. The letter identified the affected product, problem and actions to be taken. US Endoscopy requests that their customers: 1. Remove all unused affected Velocity Biopsy Valves from your inventory. 2. Complete the attached Customer Response Card and return to the attention of Kia Pennington, via fax 1-440-639-4495 or email kpenning@usendoscopy.com. 3. Mark the shipping container with code: RMA xxxxxxxx QIF, and return the product to US Endoscopy via Fed Ex Account #293799547. For questions contact Kia Pennington in Customer Service at 1-800-769-8226.
Quantity
2,173 units
Distribution
Worldwide Distribution - US Nationwide in the states of: CA, CO, IA, MA, OH, TX & WV and the countries of: Brazil, Costa Rica, Croatia, Czech Republic, Egypt, France, Germany, India, Indonesia, Italy, Latvia, Lithuania, Philippines, Poland, Romania, Russian Federation, Slovenia & South Africa.
Lot, serial or model codes
M/N: 00711140; Lot #'s: 1401267, 1404503, 1408953, 1413362, 1417521, 1420739, 1422288, 1422747, 1424014, and 1500443.
510(k) numbers
["K092461"]
PMA numbers
not on file
Product code
OCX
Firm FEI
1000122656
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register