FDA Medical Device Recalls (openFDA)
Meridian Bioscience Inc: illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed to be used with illumigene Molecular Assays. The device incubates illumigene Test Devices and reads endpoint reactions.
- Recall number
- Z-1941-2014
- Recalling firm
- Meridian Bioscience Inc
- Product
- illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed to be used with illumigene Molecular Assays. The device incubates illumigene Test Devices and reads endpoint reactions.
- Status
- Terminated
- Initiated
- 2014-04-22
- Posted
- 2014-06-27
- Terminated
- 2014-06-30
- Reason
- Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.
- Root cause
- Employee error
- Action
- Meridian sent an Urgent Medical Device Recall letter dated April 25, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately return their unit to Meridian Bioscience. Meridian's Technical Support Staff will call them directly to provide them with a return authorization number and instructions for decontaminating, packaging and shipping the unit. For additional information customers were instructed to call 800-343-3858 (North America or 513-271-3700 (Worldwide).
- Quantity
- 35 instruments
- Distribution
- Worldwide Distribution - USA including AL, AZ, CA, CO, FL, IA, IN, LA, MA, ME, MI, MN, NJ, NY, OH, PA, and WV. Internationally to Australia.
- Lot, serial or model codes
- Serial Numbers: IP2669, IP2670, IP2671, IP2672, IP2673, IP2674, IP2675, IP2676, IP2677, IP2678, IP2680, IP2682, IP2683, IP2684, IP2685, IP2686, IP2687, IP2688, IP2689, IP2690, IP2691, IP2692, IP2693, IP2694, IP2695, IP2696, IP2697, IP2698, IP2699, IP2700, IP2701, IP2702, IP2703, IP2704, IP2705, IP2706.
- 510(k) numbers
- ["K110012"]
- PMA numbers
- not on file
- Product code
- OMN
- Firm FEI
- 1524213
- Firm city
- Cincinnati
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28