FDA Medical Device Recalls (openFDA)
Ad-Tech Medical Instrument Corporation: Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bits are 30cm long and made of stainless steel. The drill kits are supplied in a sterile state and are single-use only. Used to drill cranial holes using a stereotactic frame.
- Recall number
- Z-1938-2018
- Recalling firm
- Ad-Tech Medical Instrument Corporation
- Product
- Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bits are 30cm long and made of stainless steel. The drill kits are supplied in a sterile state and are single-use only. Used to drill cranial holes using a stereotactic frame.
- Status
- Terminated
- Initiated
- 2017-12-20
- Posted
- not on file
- Terminated
- 2018-10-12
- Reason
- There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa.
- Root cause
- Packaging process control
- Action
- The firm, Ad-Tech Medical Instrument Corporation, sent a "MEDICAL DEVICE RECALL" letter on approximately 12/20/2017 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to examine inventory and quarantine any affected product on hand; notify customers of the recall if affected product was further distributed; complete and return the Acknowledgement and Receipt Form via Fax to: 262-634-5668, Attention Customer Support; Mail to: AD-TECH MEDICAL INSTRUMENT CORPORATION, 1901 WILLIAM STREET, RACINE, WI 53404 or Email to: customersupport@adtechmedical.com, and contact an Ad-Tech Customer Support Specialist for a Return Material Authorization (RMA) number. If you have any further questions, contact Regulatory Team Representative and CAPA/Complaints Manager, Monday through Friday, 7:00 AM to 5:00 PM, Central Time via FAX: 262-634-5668, Toll Free: 1-800-776-1555 or Email: customersupport@adtechmedical.com.
- Quantity
- 30 devices
- Distribution
- Worldwide Distribution: US (nationwide) to states of: CA, DC, FL, MI, MA, NC, NY, WA, WI; and country of: Canada. Distributed internationally to the United Kingdom.
- Lot, serial or model codes
- Catalog No. DDK2-2.4-30X, Lot No. 0111664 208140649, UDI No. 01008418231011741719080110011174 5208140649; Catalog No. DDK2-2.8-30X, Lot No. 0111745 208140649, UDI No. 01008418231011671719080110011166 4208140649
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 2183456
- Firm city
- Oak Creek
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as