Ortho-Clinical Diagnostics: VITROS 350 Chemistry System, Product Code 6802153, (UDI # 10758750002054) Product Usage: Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.

Recall number
Z-1936-2019
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS 350 Chemistry System, Product Code 6802153, (UDI # 10758750002054) Product Usage: Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.
Status
Terminated
Initiated
2019-04-08
Posted
not on file
Terminated
2020-11-12
Reason
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
Root cause
Device Design
Action
On 08 April 2019, a customer letter (Ref. CL2019-090) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) to either all customers who were shipped affected VITROS 250 or 350 Systems. This communication provided the following required actions: - As per the VITROS CK Slide Instructions for Use, perform Quality Control (QC) testing daily to help ensure the slides are performing within acceptable limits. - If QC results are outside of the established ranges, discard the cartridge and repeat QC testing using a fresh cartridge of VITROS CK Slides. - If multiple cartridges are loaded on the VITROS 250/350 System, daily QC testing must be performed for each cartridge prior to use. - Consider loading one cartridge of VITROS CK Slides on board your system at a time. - Complete and return the request form via fax to 1-888-557-3759. Foreign affiliates were informed by email on 08 April 2019 of the issue and instructed to notify affected consignees of the issue and appropriate actions to take.
Quantity
3889 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
All lots
510(k) numbers
["K922072"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register