Datascope Corporation: MEGA 7.5Fr. 30cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK MEGA 7.5Fr. 40cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK MEGA 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK MEGA 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit, STATLOCK & APA

Recall number
Z-1935-2016
Recalling firm
Datascope Corporation
Product
MEGA 7.5Fr. 30cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK MEGA 7.5Fr. 40cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK MEGA 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK MEGA 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit, STATLOCK & APA
Status
Terminated
Initiated
2016-05-13
Posted
not on file
Terminated
2019-03-22
Reason
It was discovered during post-quarantine packaging on 2/16/2016 that the IFU for the STATLOCK Sheath Stabilization device for Percutaneous Sheath Introducer (Vendor PN 2403097-1405R) was erroneously packaged with the Mega and Sensation Plus IAB Catheter Kits instead of the IFU for the STATLOCK Catheter Stabilization device for IAB catheters (Vendor PN 24030905-1406R).
Root cause
Unknown/Undetermined by firm
Action
Maquet Inc sent an Urgent Medical Device Field Correction Letter/Response Form dated 5/13/2016 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
Quantity
31,030 units (US); 15,604 units (OUS)
Distribution
Worldwide Distribution.
Lot, serial or model codes
(PN 0684-00-0294-01U) (PN 0684-00-0294-02U) (PN 0684-00-0295-01U) (PN 0684-00-0295-02U) (PN 0684-00-0296-01U)
510(k) numbers
["K120868"]
PMA numbers
not on file
Product code
DSP
Firm FEI
1000222374
Firm city
Fairfield
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register