Zimmer, Inc.: PathFinder NXT Pivoting Percutaneous Rod Holder This device is an instrument used during orthopedic surgery. The intended use of the pivoting percutaneous rod holder is to attach to the rod to facilitate rod insertion, positioning, and orientation through the extender sleeves, under the fascia. The rod is secured to the holder by tightening a captive set screw against the rod end in the tip of the holder using the set screw driver. After positioning the rod and securing at least one pedicle screw closure top against the implanted rod, the holder is removed from the rod end by loosening the captive set screw with the set screw driver.

Recall number
Z-1933-2012
Recalling firm
Zimmer, Inc.
Product
PathFinder NXT Pivoting Percutaneous Rod Holder This device is an instrument used during orthopedic surgery. The intended use of the pivoting percutaneous rod holder is to attach to the rod to facilitate rod insertion, positioning, and orientation through the extender sleeves, under the fascia. The rod is secured to the holder by tightening a captive set screw against the rod end in the tip of the holder using the set screw driver. After positioning the rod and securing at least one pedicle screw closure top against the implanted rod, the holder is removed from the rod end by loosening the captive set screw with the set screw driver.
Status
Terminated
Initiated
2012-06-05
Posted
2012-07-03
Terminated
2013-02-14
Reason
Complaints have been reported where, during surgery, the surgeon was unable to loosen or remove the inserted rod from the rod holder instrument. The complaint investigation showed the hex mating feature of the set screw was broken and material missing. There were no reports of fragments of the set screw being left in the patient or that there was any patient or user injury, but these events we
Root cause
Device Design
Action
The firm, Zimmer Spine, sent an "URGENT MEDICAL DEVICE RECALL" letter dated June 5, 2012 to all consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to review the notification and ensure they are aware of the content; to immediately remove and discontinue use of Pivoting Percutaneous Rod Holder (part number 3564-1) from Pathfinder NXT instrument sets in their possession and follow the PathFinder NXT Surgical Technique; to notify their Zimmer Spine customer service representative to obtain return authorization; and return the complete Surgeon Contact Certification Form within 10 working days from receipt of notification to ronald.musselman@zimmer.com or Fax to 1-512-258-0995. If you have any questions, call Zimmer Spine Customer Service at 866-774-6368.
Quantity
77 units
Distribution
Worldwide distribution: USA (nationwide) and countries of: Australia, France and Germany.
Lot, serial or model codes
Part number 3564-1; all lots
510(k) numbers
["K100845"]
PMA numbers
not on file
Product code
NKB
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register