FDA Medical Device Recalls (openFDA)
Boston Scientific CRM Corp: Boston Scientific TELIGEN, Implantable Cardiac Defibrillator, Models E102 and E110. Sterilized using ethylene oxide. Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 66112-5798, USA. The product is indicated for patients who are at risk for sudden cardiac death due to ventricular arrhythmias and may require pacing support. COGNIS devices also provide bi-ventricular pacing.
- Recall number
- Z-1933-2009
- Recalling firm
- Boston Scientific CRM Corp
- Product
- Boston Scientific TELIGEN, Implantable Cardiac Defibrillator, Models E102 and E110. Sterilized using ethylene oxide. Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 66112-5798, USA. The product is indicated for patients who are at risk for sudden cardiac death due to ventricular arrhythmias and may require pacing support. COGNIS devices also provide bi-ventricular pacing.
- Status
- Terminated
- Initiated
- 2009-07-20
- Posted
- 2009-08-18
- Terminated
- 2012-06-26
- Reason
- Boston Scientific Cardiac Rhythm Management retrieved devices that could be subject to a potential for acute non-secure lead connections when implanted. Nonsecure lead connections can manifest as high impedance, or lead electrogram artifacts resulting in oversensing and/or inappropriate delivery of therapy.
- Root cause
- Device Design
- Action
- Boston Scientific Corporation field personnel began retrieving devices containing the original set screw system on Friday, July 17, 2009. If needed a letter, dated 07/17/2009 was left with the hospital administrator letting them know that product has been collected for return to Boston Scientific Corporation. Replacement devices will be provided at no charge. If the field personnel was unable to contact the hospital in the time provided, a notification letter dated July 22, 2009 was sent via Federal Express to the hospital. This letter was directed to the EP Catheter Lab Manager with a cc to the Hospital Risk Manager, Hospital Administrator and Purchasing and Central Supply. The letter asked that the affected devices not be implanted and to contact their local representative for pick-up and return. The firm will replace the affected devices. Please direct questions to your local Sales representative or Boston Scientific Corporation CRM Technical Services by calling 1-800-227-3422.
- Quantity
- 104
- Distribution
- U.S. Distribution only (states of AZ, CA, CT, FL, GA, IA, IL, KY, MD, MO, NH, NJ, NY, OH, PA, SC, TN, TX, WI, and Washington D.C.) and Puerto Rico.
- Lot, serial or model codes
- Model E102, Serial numbers: 002673, 003415, 004433, 004856, 005108, 005677, 006215, 006309, 006321, 006641, 006766, 006771, 006787, 006799, 006815, 006819, 006937, 006946, 006960, 006969, 006970, 006979, 006985, 007011, 007051, 007125, 007219, 007706, 007713, 007730, 008173, 008459, 008541, 008617, 008759, 008822, 008832, 008972, 008998, 009055, 009068, 009079, 009102, 009117, 009151, 009182, 009224, 009345, 009379, 009723, 010141, 010167 and 010186 and Model E110, Serial numbers: 002872, 003458, 003681, 004229, 004300, 004979, 004989, 006517, 006935, 007106, 008280, 008356, 008591, 009837, 009865, 010026, 010341, 010365, 010393, 010413, 010508, 010510, 010533, 010555, 010581, 010596, 010694, 011538, 011578, 012363, 012373, 012571, 012712, 012785, 012809, 013015, 013024, 013759, 013885, 111500, 111570, 111584, 111599, 111610, 111661, 111664, 111700, 111703, 112637, 112716 and 113694.
- 510(k) numbers
- not on file
- PMA numbers
- ["P010012S165","P960040S155"]
- Product code
- LWP
- Firm FEI
- 2124215
- Firm city
- Saint Paul
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28