Skip to the content

FDA Medical Device Recalls (openFDA)

SpineFrontier, Inc.: T-Handle, Strike Plate, Push Connection, Part Number: SI50023 Rev A T-handles are silicone handled instruments that have a quick release end. The silicone T-handle has an impactor cap on the proximal end of the handle which is designed to be struck with a mallet. T-handles are used primarily to attach to, and operate a wide range of instruments used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients.

Recall number
Z-1932-2013
Recalling firm
SpineFrontier, Inc.
Product
T-Handle, Strike Plate, Push Connection, Part Number: SI50023 Rev A T-handles are silicone handled instruments that have a quick release end. The silicone T-handle has an impactor cap on the proximal end of the handle which is designed to be struck with a mallet. T-handles are used primarily to attach to, and operate a wide range of instruments used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients.
Status
Terminated
Initiated
2012-08-29
Posted
2013-08-08
Terminated
2013-10-18
Reason
During an operation the strike plate of a T-Handle loosened and separated from the T-Handle.
Root cause
Process design
Action
On August 29, 2011, The Supply Chain Team was notified of the recall via email. They were informed that all handles are being pulled from the field. This resulted in the return of all T-Handles to SpineFrontier. Further questions please call (978) 232-3990.
Quantity
11
Distribution
US Distribution including the states of MA, NV and TX.
Lot, serial or model codes
Lot Numbers: 1335201A, 1446101A
510(k) numbers
["K092815"]
PMA numbers
not on file
Product code
MDM
Firm FEI
3005977257
Firm city
Beverly
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register