Skip to the content

FDA Medical Device Recalls (openFDA)

MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK: MHI-TM2000 Linear Accelerator System, Software Version 3.0.0 revision 13 or earlier; Proprietary and Trade Name: MHI-TM2000/ VERO (TM); Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated

Recall number
Z-1931-2013
Recalling firm
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
Product
MHI-TM2000 Linear Accelerator System, Software Version 3.0.0 revision 13 or earlier; Proprietary and Trade Name: MHI-TM2000/ VERO (TM); Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated
Status
Terminated
Initiated
2013-06-21
Posted
2013-08-08
Terminated
2017-03-23
Reason
Software anomaly: Mitsubishi Heavy Industries, Ltd. became aware of a potential safety issue associated with the VERO (TM) Linear Accelerator System. The VERO (TM) may save treatment delivery record into wrong location under certain conditions.
Root cause
Software Design Change
Action
The firm, Mitsubishi Heavy Industries, Ltd. (MHI), sent an "URGENT FIELD SAFETY NOTICE" letter dated June 21, 2013 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: If any minor failure occurs, please always press [ALARM RESET] button on operator console first; and if [EXIT] button is clicked after a minor failure occurs, please click [YES] on the pop up dialog to finish treatment completely. MHI is developing a correction for this issue. Tentatively planned availability date: End of October 2013. The correction will be free of charge and a MHI Customer Service representative will contact affected customers to schedule its installation as the correction becomes available. If you need any further information or support concerning this issue, please feel free to contact our Customer Service representative: VERO GmbH, Kapellenstrasse 12, 8566 Feldkirchen Germany, at +49-89-99-1568-44 or email: franz.gum@vero-sbrt.com.
Quantity
14 (2 in US)
Distribution
Worldwide distribution: US (nationwide) including states of TX and FL; and countries of: Belgium, Italy, Japan and Korea.
Lot, serial or model codes
Serial Numbers #s 201902 and 203901
510(k) numbers
["K072047","K122450"]
PMA numbers
not on file
Product code
IYE
Firm FEI
3006942329
Firm city
HIROSHIMA
Firm state
4-CHOME KAN-ON-SHIN-MACHI
Firm country
Japan

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register