Abiomed, Inc.: iPulse Circulatory Support System Catalog number: 0036-0010 The iPulse System Console is a multi-purpose mechanical circulatory support system used to support patients suffering from acute cardiovascular failure. The iPulse console can be used to drive exclusively one of two classes of therapeutic devices: either pneumatically actuated Ventricular Assist Devices (VADs) for total circulatory support, or (i.e., non-simultaneously) an intra-aortic balloon (IAB) catheter for counter-pulsation therapy.

Recall number
Z-1931-2009
Recalling firm
Abiomed, Inc.
Product
iPulse Circulatory Support System Catalog number: 0036-0010 The iPulse System Console is a multi-purpose mechanical circulatory support system used to support patients suffering from acute cardiovascular failure. The iPulse console can be used to drive exclusively one of two classes of therapeutic devices: either pneumatically actuated Ventricular Assist Devices (VADs) for total circulatory support, or (i.e., non-simultaneously) an intra-aortic balloon (IAB) catheter for counter-pulsation therapy.
Status
Terminated
Initiated
2009-04-23
Posted
2009-09-16
Terminated
2014-06-05
Reason
iPulse Console SC1035 displayed a "Low Pressure / Low Flow" alarm and stopped pumping during patient transport in battery operation.
Root cause
Device Design
Action
An "Urgent Voluntary Medical Device Field Correction Notice" was issued April 8, 2009 to affected customers. The customers were notified that the iPulse Console SC1032 displayed a "Low Pressure/ Low Flow" alarm and stopped pumping during patient transport in battery operation. Instructions were provided to users in case they encounter this failure mode. ABIOMED Field Service Engineers will contact customers to schedule a time to visit site and correct affected consoles. Please contact ABIOMED's Field Service at 1.800.554.8666, if you have any questions.
Quantity
79
Distribution
Domestic: MA, MI, SC, NY, NJ, OH, CA, PA, AR, MI, MD, CT, IL, WI, TX, NJ, TN, IN, MI, DE, FL, IA, WA, KY, and WV Foreign: Germany, Greece, Italy, Latvia, Spain, and Turkey
Lot, serial or model codes
Serial numbers: SC1000, SC1001, SC1003 - SC1010, SC1014 - SC1023, SC1025 - SC1070, SC1072 - SC1081, SC1083, SC1084, and SC1089
510(k) numbers
not on file
PMA numbers
["P900023"]
Product code
DSQ
Firm FEI
1220648
Firm city
Danvers
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register