FDA Medical Device Recalls (openFDA)
Abiomed, Inc.: iPulse Circulatory Support System Catalog number: 0036-0010 The iPulse System Console is a multi-purpose mechanical circulatory support system used to support patients suffering from acute cardiovascular failure. The iPulse console can be used to drive exclusively one of two classes of therapeutic devices: either pneumatically actuated Ventricular Assist Devices (VADs) for total circulatory support, or (i.e., non-simultaneously) an intra-aortic balloon (IAB) catheter for counter-pulsation therapy.
- Recall number
- Z-1931-2009
- Recalling firm
- Abiomed, Inc.
- Product
- iPulse Circulatory Support System Catalog number: 0036-0010 The iPulse System Console is a multi-purpose mechanical circulatory support system used to support patients suffering from acute cardiovascular failure. The iPulse console can be used to drive exclusively one of two classes of therapeutic devices: either pneumatically actuated Ventricular Assist Devices (VADs) for total circulatory support, or (i.e., non-simultaneously) an intra-aortic balloon (IAB) catheter for counter-pulsation therapy.
- Status
- Terminated
- Initiated
- 2009-04-23
- Posted
- 2009-09-16
- Terminated
- 2014-06-05
- Reason
- iPulse Console SC1035 displayed a "Low Pressure / Low Flow" alarm and stopped pumping during patient transport in battery operation.
- Root cause
- Device Design
- Action
- An "Urgent Voluntary Medical Device Field Correction Notice" was issued April 8, 2009 to affected customers. The customers were notified that the iPulse Console SC1032 displayed a "Low Pressure/ Low Flow" alarm and stopped pumping during patient transport in battery operation. Instructions were provided to users in case they encounter this failure mode. ABIOMED Field Service Engineers will contact customers to schedule a time to visit site and correct affected consoles. Please contact ABIOMED's Field Service at 1.800.554.8666, if you have any questions.
- Quantity
- 79
- Distribution
- Domestic: MA, MI, SC, NY, NJ, OH, CA, PA, AR, MI, MD, CT, IL, WI, TX, NJ, TN, IN, MI, DE, FL, IA, WA, KY, and WV Foreign: Germany, Greece, Italy, Latvia, Spain, and Turkey
- Lot, serial or model codes
- Serial numbers: SC1000, SC1001, SC1003 - SC1010, SC1014 - SC1023, SC1025 - SC1070, SC1072 - SC1081, SC1083, SC1084, and SC1089
- 510(k) numbers
- not on file
- PMA numbers
- ["P900023"]
- Product code
- DSQ
- Firm FEI
- 1220648
- Firm city
- Danvers
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28