Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics: VITROS Chemistry Products CKMB Slides, Creatine Kinase MB, Product Code 805 8232-300 slides, UDI Number 10758750004294 - Product Usage: The VITROS Chemistry Product Slides are used to measure creatine kinase MB (CK-MB) activity in serum which used to be ordered to detect Myocardial Infarction.

Recall number
Z-1930-2020
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS Chemistry Products CKMB Slides, Creatine Kinase MB, Product Code 805 8232-300 slides, UDI Number 10758750004294 - Product Usage: The VITROS Chemistry Product Slides are used to measure creatine kinase MB (CK-MB) activity in serum which used to be ordered to detect Myocardial Infarction.
Status
Terminated
Initiated
2020-04-01
Posted
not on file
Terminated
2021-01-19
Reason
The Spanish and Swedish language versions of the IFU state that the assay can be tested with Plasma which is false and could result in biased results.
Root cause
Labeling design
Action
An Urgent Product Correction letter was sent on April 1, 2020 to affected customers stating that the Spanish and Swedish IFU versions for the VITROS CK-MB slides incorrectly included plasma as a sample type in the Intended Use section. Customers were instructed to complete and return the Confirmation of Receipt form then, determine whether the effected IFUs were being used in their laboratory and identify any plasma samples that were taken while using those instructions for use. The laboratory directors should determine how to handle the assays that used plasma samples. The firm directed customers to go to the Ortho-Clinical Diagnostics website to download the updated Spanish and Swedish IFUs.
Quantity
not on file
Distribution
US Nationwide distribution.
Lot, serial or model codes
CKMB Slides that expire after July 20, 2018 and within current expiry.
510(k) numbers
["K861387"]
PMA numbers
not on file
Product code
JHS
Firm FEI
1000305840
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register