Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Dimension Flex Reagent Cartridge C-Reactive Protein Extended Range (RCRP). In vitro diagnostic test intended for the quantitative determination of CRP in human serum and plasma (lithium heparin). Measurement of C-Reactive Protein is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases.

Recall number
Z-1930-2009
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Dimension Flex Reagent Cartridge C-Reactive Protein Extended Range (RCRP). In vitro diagnostic test intended for the quantitative determination of CRP in human serum and plasma (lithium heparin). Measurement of C-Reactive Protein is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases.
Status
Terminated
Initiated
2008-04-15
Posted
2009-09-08
Terminated
2009-09-09
Reason
falsely elevated results in EDTA plasma samples
Root cause
Labeling False and Misleading
Action
Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice on April 15, 2008 informing their customers that the EDTA plasma specimen claim had been rescinded. Current lots in the firm's inventory have been put on hold and reworked with an Alert Card rescinding the EDTA plasma claim. Subsequent lots use a revised instruction for use stating "Do not use EDTA plasma" in the Specimen Collection and Handling Section. For further questions, contact Siemens Healthcare Diagnostics at 1-302-631-6564.
Quantity
25,760 cartons
Distribution
Worldwide Distribution -- US, Austria, Bosnia, Czech Republic, Hungary, Slovakia, Serbia, Denmark, Estonia, Latvia, Lithuania, Norway, Sweden, Belgium, Cyprus, Canary Islands, France, Great Britain, Germany, Ireland, Italy, Netherlands, Poland, Russia, Portugal, Saudi Arabia, Spain, South Africa, Switzerland, Turkey, Yemen, Bahamas, Brazil, Canada, Mexico, Australia, Bangladesh, China, Hong Kong, Japan, Malaysia, Pakistan, and Taiwan.
Lot, serial or model codes
Catalog Number DF34; Lot Numbers: CE8324 exp 11/19/2008, BA8349 exp 12/14/2008, FB9003 exp1/3/2009, BC9017 exp1/17/2009, EC9021 exp1/21/2009, DB9032 exp2/1/2009, and DC9045 exp2/14/2009.
510(k) numbers
["K003419"]
PMA numbers
not on file
Product code
DCN
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register