FDA Medical Device Recalls (openFDA)

Encore Medical, LP: EMPOWR Acetabular System, Non-Sterile, Model Numbers 803-15-023 803-15-026 803-15-027 803-15-028 803-15-029 803-15-030 Product Usage: The EMPOWR Acetabular Impactor Balls is intended for use in joint replacement in patients suffering from disability due to noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; rheumatoid arthritis, correction of functional deformity, femoral fracture, and in the salvage of previously failed surgical attempts.

Recall number
Z-1929-2020
Recalling firm
Encore Medical, LP
Product
EMPOWR Acetabular System, Non-Sterile, Model Numbers 803-15-023 803-15-026 803-15-027 803-15-028 803-15-029 803-15-030 Product Usage: The EMPOWR Acetabular Impactor Balls is intended for use in joint replacement in patients suffering from disability due to noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; rheumatoid arthritis, correction of functional deformity, femoral fracture, and in the salvage of previously failed surgical attempts.
Status
Open, Classified
Initiated
2019-10-22
Posted
not on file
Terminated
not on file
Reason
The firm received complaints that the EMPOWR Acetabular Impactor Balls were fracturing/cracking during impaction which could cause debris to migrate or be left in the joint space.
Root cause
Device Design
Action
The firm requested its agents to identify and return affected devices.
Quantity
86 units
Distribution
US Nationwide distribution including in the states of MN, VA, IL, IN, SC, N, TX, CA.
Lot, serial or model codes
All lots
510(k) numbers
["K190057"]
PMA numbers
not on file
Product code
LPH
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register