Edwards Lifesciences, LLC: Oxymetry Cable, Rxonly. REF/UDI: HEMOXSC100/(01)00690103196986 - Product Usage: The HemoSphere oximetry cable is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure venous oxygen saturation.

Recall number
Z-1928-2020
Recalling firm
Edwards Lifesciences, LLC
Product
Oxymetry Cable, Rxonly. REF/UDI: HEMOXSC100/(01)00690103196986 - Product Usage: The HemoSphere oximetry cable is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure venous oxygen saturation.
Status
Terminated
Initiated
2019-10-07
Posted
2020-05-08
Terminated
2021-04-27
Reason
Intermittent communication due to inconsistent crimping of connectors onto cables.
Root cause
Process control
Action
On xx/xx/2019, and URGENT MEDICAL DEVICE RECALL-ACTION REQUIRED, letter was sent to customers by FedEx. Customers are instructed to acknowledge this notification , contact the firm at the contact information in the communication to obtain a Returned Goods Authorization (RGA) number and to return all product to recalling firm.
Quantity
35
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, CA, FL, IL, IN and MO, and the countries of Germany, Netherlands, United Kingdom.
Lot, serial or model codes
The cables with the following serial numbers (35) are recalled: 13710131, 13710130, 13710129, 13710115, 13710113, 13710025, 13710027, 13710028, 13710029, 13710030, 13710031, 13710041, 13710043, 13710045, 13710050, 13710051, 13710056, 13710057, 13710058, 13710061, 13710062, 13710063, 13710064, 13710066, 13710069, 13710070, 13710084, 13710086, 13710089, 13710092, 13710093, 13710094, 13710098, 13710099, 13710100.
510(k) numbers
["K163381"]
PMA numbers
not on file
Product code
DQK
Firm FEI
2015691
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register