Skip to the content

FDA Medical Device Recalls (openFDA)

Hamilton Medical Inc: Hamilton-G5 Ventilator with software version V2.00x and Neonatal Option or V2.01x and Neonatal software. Hamilton-G5: catalog number 159001; Neonatal Software catalog number: 159700; Neonatal Option catalog number: 159187. Hamilton Medical Intended for use in the hospital and institutional environment where healthcare professionals provide patient care.

Recall number
Z-1928-2011
Recalling firm
Hamilton Medical Inc
Product
Hamilton-G5 Ventilator with software version V2.00x and Neonatal Option or V2.01x and Neonatal software. Hamilton-G5: catalog number 159001; Neonatal Software catalog number: 159700; Neonatal Option catalog number: 159187. Hamilton Medical Intended for use in the hospital and institutional environment where healthcare professionals provide patient care.
Status
Terminated
Initiated
2011-02-16
Posted
2011-04-08
Terminated
2012-03-21
Reason
The adaptive volume/pressure regulation can be automatically and permanently reduced to PEEP + 3 mbar when a closed suctioning maneuver is performed on the patient and may be undetected by the operator.
Root cause
Software design
Action
A Medical device Safety Alert and Corrective Action along with a cover letter, was sent to all locations on February 16, 2011. The letter identified the affected product along with the reason for recall. The letter states that operators deactivate the "high tidal volume" alarm as described in appendix K.4 of the operator's manual until the ventilator is upgraded with a corrected software. Distributors will upgrade the affected devices with the corrected software as soon as possible. Questions can be directed to 800-426-6331, ext 215.
Quantity
97 units
Distribution
Nationwide Distribution -- MA, FL, GA, NC, MI, NC, MO, TX, and WV.
Lot, serial or model codes
Software version V2.00x and Neonatal Option or V2.01x and Neonatal software
510(k) numbers
["K081521"]
PMA numbers
not on file
Product code
CBK
Firm FEI
2937708
Firm city
Reno
Firm state
NV
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register