Skip to the content

FDA Medical Device Recalls (openFDA)

ICU Medical, Inc.: The Cogent Hemodynamic Monitoring System. REF58400-000, GTIN 00840619099459; REF 58400-000R, GTIN 0087709094314; REF 58403, GTIN 00840619079505; REF 58401, GTIN 00840619079499 - Product Usage: intended for patients for whom the monitoring of continuous cardiac output and calculated hemodynamic parameters is indicated for diagnostic and prognostic evaluation by a clinician. The target population includes Critical Care Patients, Trauma Patients, and Cardiac Surgery Patients.

Recall number
Z-1927-2020
Recalling firm
ICU Medical, Inc.
Product
The Cogent Hemodynamic Monitoring System. REF58400-000, GTIN 00840619099459; REF 58400-000R, GTIN 0087709094314; REF 58403, GTIN 00840619079505; REF 58401, GTIN 00840619079499 - Product Usage: intended for patients for whom the monitoring of continuous cardiac output and calculated hemodynamic parameters is indicated for diagnostic and prognostic evaluation by a clinician. The target population includes Critical Care Patients, Trauma Patients, and Cardiac Surgery Patients.
Status
Terminated
Initiated
2020-02-14
Posted
not on file
Terminated
2023-05-03
Reason
Firm identified software issues which leads to the patient ID and patient information not being stored in the system.
Root cause
Software Design Change
Action
On 14 February 2020, the firm sent notification of the recall to direct consignees of record. Delivery will be confirmed via traceable mailing. All consignees who further distributed the product were asked to notify their customers
Quantity
123 units
Distribution
US Nationwide distribution including in the states of AR, CA, FL, IL, MD, NC, NE, NY, TX, VT, WA and WI.
Lot, serial or model codes
All serial numbers of Cogent HMS, List Number(s): 58400-000; 58400-000R; 58403-000; 58401-000 with Software Versions 1.3.3.42 and prior are impacted.
510(k) numbers
["K152006"]
PMA numbers
not on file
Product code
DXG
Firm FEI
2025816
Firm city
San Clemente
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register