FDA Medical Device Recalls (openFDA)
ICU Medical, Inc.: The Cogent Hemodynamic Monitoring System. REF58400-000, GTIN 00840619099459; REF 58400-000R, GTIN 0087709094314; REF 58403, GTIN 00840619079505; REF 58401, GTIN 00840619079499 - Product Usage: intended for patients for whom the monitoring of continuous cardiac output and calculated hemodynamic parameters is indicated for diagnostic and prognostic evaluation by a clinician. The target population includes Critical Care Patients, Trauma Patients, and Cardiac Surgery Patients.
- Recall number
- Z-1927-2020
- Recalling firm
- ICU Medical, Inc.
- Product
- The Cogent Hemodynamic Monitoring System. REF58400-000, GTIN 00840619099459; REF 58400-000R, GTIN 0087709094314; REF 58403, GTIN 00840619079505; REF 58401, GTIN 00840619079499 - Product Usage: intended for patients for whom the monitoring of continuous cardiac output and calculated hemodynamic parameters is indicated for diagnostic and prognostic evaluation by a clinician. The target population includes Critical Care Patients, Trauma Patients, and Cardiac Surgery Patients.
- Status
- Terminated
- Initiated
- 2020-02-14
- Posted
- not on file
- Terminated
- 2023-05-03
- Reason
- Firm identified software issues which leads to the patient ID and patient information not being stored in the system.
- Root cause
- Software Design Change
- Action
- On 14 February 2020, the firm sent notification of the recall to direct consignees of record. Delivery will be confirmed via traceable mailing. All consignees who further distributed the product were asked to notify their customers
- Quantity
- 123 units
- Distribution
- US Nationwide distribution including in the states of AR, CA, FL, IL, MD, NC, NE, NY, TX, VT, WA and WI.
- Lot, serial or model codes
- All serial numbers of Cogent HMS, List Number(s): 58400-000; 58400-000R; 58403-000; 58401-000 with Software Versions 1.3.3.42 and prior are impacted.
- 510(k) numbers
- ["K152006"]
- PMA numbers
- not on file
- Product code
- DXG
- Firm FEI
- 2025816
- Firm city
- San Clemente
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28