Siemens Healthcare Diagnostics, Inc: Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot number: 454778 Kit Catalog Number: 112219 SMN: 10310026 Intended Use: In vitro diagnostic use. Acid Regent is a component of the ADVIA Centaur and ADVIA Centaur XP instruments
- Recall number
- Z-1927-2014
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Product
- Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot number: 454778 Kit Catalog Number: 112219 SMN: 10310026 Intended Use: In vitro diagnostic use. Acid Regent is a component of the ADVIA Centaur and ADVIA Centaur XP instruments
- Status
- Terminated
- Initiated
- 2014-06-02
- Posted
- 2014-06-26
- Terminated
- 2015-02-05
- Reason
- Swollen acid bottles due to a trace metal contamination in the Acid Reagent bottles
- Root cause
- Material/Component Contamination
- Action
- Siemens Healthcare Diagnostics issued Urgent Medical Device Recall (UMDR) to customers in the United States on June 02, 2014 and a Urgent Field Safety Notice (UFSN) issued outside the United States on June 02, 2014. These notices inform the customer of the potential for swollen acid bottles due to confirmed metal contamination. Customers with alternative lots in inventory were instructed to transition to the alternative lot upon receipt of the communication. Customers without an alternative lot of acid in inventory were informed that Siemens proactively shipped an alternative lot to their laboratory and they should transition to that lot upon its receipt. Siemens also informed customers without an alternative lot that they may continue to run and report patient results using the affected lot of acid. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
- Quantity
- 3,839 kits
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Canada, Argentina, China, Colombia, Korea, Malyasia, New Zealand, Mexico, Guatemala, and Singapore.
- Lot, serial or model codes
- Kit Lot Number: 454779 Exp. Date: 28 February, 2015
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 1219913
- Firm city
- East Walpole
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28