Skip to the content

FDA Medical Device Recalls (openFDA)

SpineFrontier, Inc.: Rev E PedFuse Return Screw Inserter, Part Number: 11-80030 Screw inserters are instruments which capture, hold and drive pedicle screw assemblies. There is a capturing feature that engages and holds the screw in place. There is also a feature used to lock the screw to the inserter before insertion of the screw. They can also be used to remove screws or back screws out. Screw inserters are used during the implantation of spinal pedicle screws intended to provide immobilization and stabilization of spinal segments in skeletally mature patients. Used in conjunction with Pedicle Screw Spinal system marketed as KRD1 PedFuse Pedicle Screw System .

Recall number
Z-1926-2013
Recalling firm
SpineFrontier, Inc.
Product
Rev E PedFuse Return Screw Inserter, Part Number: 11-80030 Screw inserters are instruments which capture, hold and drive pedicle screw assemblies. There is a capturing feature that engages and holds the screw in place. There is also a feature used to lock the screw to the inserter before insertion of the screw. They can also be used to remove screws or back screws out. Screw inserters are used during the implantation of spinal pedicle screws intended to provide immobilization and stabilization of spinal segments in skeletally mature patients. Used in conjunction with Pedicle Screw Spinal system marketed as KRD1 PedFuse Pedicle Screw System .
Status
Terminated
Initiated
2013-03-14
Posted
2013-08-08
Terminated
2015-06-03
Reason
Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the inserter which allows the collar to slide off the proximal end of the inserter. When the collar is not in place the locking sleeve can also slide off the inserter.
Root cause
Device Design
Action
Spine Frontier sent an email dated December 10, 2012, to the affected customers that a Pedfuse Inserter with part number PM3979 needed to be returned. Further questions please call (978) 232-3990.
Quantity
8
Distribution
Distribution only in Texas.
Lot, serial or model codes
Lot Number: PM3979
510(k) numbers
["K092420"]
PMA numbers
not on file
Product code
HXX
Firm FEI
3005977257
Firm city
Beverly
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register