Skip to the content

FDA Medical Device Recalls (openFDA)

Natus Neurology DBA Excel Tech., Ltd. (XLTEK): Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box) Model/Catalog Number: OBM00002 OBM DAB (Digital Acquisition Box) Software Version: N/A Product Description: OBM00002 OBM DAB (Digital Acquisition Box) Component: No

Recall number
Z-1925-2025
Recalling firm
Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
Product
Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box) Model/Catalog Number: OBM00002 OBM DAB (Digital Acquisition Box) Software Version: N/A Product Description: OBM00002 OBM DAB (Digital Acquisition Box) Component: No
Status
Open, Classified
Initiated
2025-05-07
Posted
2025-06-09
Terminated
not on file
Reason
Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device.
Root cause
Employee error
Action
On May 7, 2025, the firm began notifying customers through "Urgent Medical Device Recall" letters. Customers are instructed to check their Olympic Brainz Monitor stock and determine if they have any devices with the incorrect label orientation. Immediately cease use of incorrectly labeled product and quarantine until you receive return instructions from Natus. Customers with incorrect product labelling will be provided with replacement device(s). If you have any questions regarding this notice, please contact FCA@Natus.com.
Quantity
3069
Distribution
Worldwide distribution - US Nationwide including Puerto Rico (PR) and OUS (foreign) countries of: Italy (IT), Estonia (EE), Israel (IL), China (CN), Germany (DE), Saudi Arabia (SA), Switzerland (CH), Chile (CL), Colombia (CO), Netherlands (the) (NL), Japan (JP), United Kingdom of Great Britain and Northern Ireland(the) (GB), Russian Federation (the)(RU), Bolivia (Plurinational State of) (BO), Canada (CA), France (FR), Peru (PE), Guyana (GY), Mexico (MX), Belgium (BE), Thailand (TH), Paraguay (PY), Hong Kong (HK), United Arab Emirates (AE), Austria (AT), India (IN), Indonesia (ID), Kuwait (KW), Qatar (QA), Lithuania (LT), Denmark (DK), Myanmar (MM), Uruguay (UY), Tunisia (TN), Turkey (TR), Korea (the Republic of) (KR), Hungary (HU), Slovenia (SI), Argentina (AR), Norway (NO), Oman (OM), Poland (PL), Finland (FI), Ireland (IE), Australia (AU), Latvia (LV), Ecuador (EC), Singapore (SG), Spain (ES), Serbia (RS), South Africa (ZA), Romania (RO), Algeria (DZ), Portugal (PT), Dominican Republic (the) (DO), Lebanon (LB), Malaysia (MY), Costa Rica (CR), Viet Nam (VN), & Bahrain (BH).
Lot, serial or model codes
UDI-DI 00382830010825 Serial Numbers 001 through OBM00002H3613. Note earlier version serial numbers did not contain the sequence of "OBM"
510(k) numbers
["K123079"]
PMA numbers
not on file
Product code
OMA
Firm FEI
3002809369
Firm city
Oakville
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register