Skip to the content

FDA Medical Device Recalls (openFDA)

Alcon Research LLC: ClearCut S Safety Sideport Knife 1.2mm Dual Bevel Model/Catalog Number: 8065771541 Software Version: N/A Product Description: Ophthalmic Knife Component: N/A

Recall number
Z-1925-2024
Recalling firm
Alcon Research LLC
Product
ClearCut S Safety Sideport Knife 1.2mm Dual Bevel Model/Catalog Number: 8065771541 Software Version: N/A Product Description: Ophthalmic Knife Component: N/A
Status
Open, Classified
Initiated
2024-04-22
Posted
2024-05-30
Terminated
not on file
Reason
Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Corneal/Scleral V-Lance Ophthalmic knives.
Root cause
Process control
Action
On April 22, 2024 VOLUNTARY MEDICAL DEVICE FIELD CORRECTION letters to customers. Actions to be taken by the Customer / User: If a knife does not perform as expected, or if you feel it does not meet your expectation for sharpness, discontinue use of that specific knife and replace the knife with a new one to avoid potential complications. To acknowledge your receipt of this Voluntary Medical Device Correction notification, please take the following steps: 1. Forward this notification to all departments or organizations using Alcon Ophthalmic Knives. 2. Follow the risk mitigation precautions provided in this notice when using identified catalogue numbers of ophthalmic knives. 3. Please complete the attached Response Form indicating your understanding of the included instructions and return the attached Response Form via email or fax to Alcon. Email: Market.Actions@alcon.com Fax: 817-302-4337 Alcon has reported this issue to Health Authorities in accordance with applicable regulations. In the event you have experienced adverse events or product quality issues related to this communication, please contact Alcon via web (https://notifeye.alcon.com), by email (msus.safety@alcon.com) or by phone (1-800-757-9780). Adverse events or quality problems experienced with the use of this product may also be reported to the FDA MedWatch Adverse Event Reporting program either online (FDA EMAIL), by regular mail or by fax: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm. Should you have any questions or concerns about this matter or need help finding a replacement or substitute sterile standalone ophthalmic surgical knife, please call Alcon Customer Service or contact your Alcon Sales Representative.
Quantity
4,930 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Columbia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Korea, Kuwait, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom. Product distributed as sterile stand-alone knives have been placed on HOLD. Product included within Alcon Custom Pak (surgical procedure pack) are continuing to be distributed as ceasing distribution would cause immediate market shortage and result in canceled surgeries.
Lot, serial or model codes
Model/Catalog Number: 8065771541; UDI/DI: 00380657715411; Lot/Batch: 168NME, 168NND, 16AV17;
510(k) numbers
["K110166"]
PMA numbers
not on file
Product code
HNN
Firm FEI
1610287
Firm city
Fort Worth
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register