Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: DISCOVERY ELBOW prosthesis, 4 MM X 100 MM RIGHT HUMERAL FLANGED I BOND COAT Tl 6AL 4V ALLOY. Elbow joint replacement prostheses are intended for primary and revision joint arthroplasty for use in cemented applications.

Recall number
Z-1925-2014
Recalling firm
Biomet, Inc.
Product
DISCOVERY ELBOW prosthesis, 4 MM X 100 MM RIGHT HUMERAL FLANGED I BOND COAT Tl 6AL 4V ALLOY. Elbow joint replacement prostheses are intended for primary and revision joint arthroplasty for use in cemented applications.
Status
Terminated
Initiated
2014-05-15
Posted
2014-06-26
Terminated
2015-03-16
Reason
Surface finish is different than specified. The implant has 30 grit blast on the exterior of the flange when the print calls for the glass bead blast. The patient may experience pain requiring surgical intervention.
Root cause
Nonconforming Material/Component
Action
Biomet sent an Urgent Medical Device Recall Notice letter dated May 28, 2014 to their customer. The letter identified the affected product, problem and actions to be taken. The letter instructed physician to monitor the patient for any pain that may require surgical intervention. For questions call (574) 372-1570.
Quantity
1
Distribution
US Nationwide Distribution in the state of California
Lot, serial or model codes
Catalog/REF #: 114905, Lot #: M846640
510(k) numbers
["K013042"]
PMA numbers
not on file
Product code
JDC
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register