Siemens Healthcare Diagnostics, Inc.: Dimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (DC107) Product Usage: The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of tacrolimus therapy in kidney and liver transplant patients.
- Recall number
- Z-1924-2013
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- Dimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (DC107) Product Usage: The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of tacrolimus therapy in kidney and liver transplant patients.
- Status
- Terminated
- Initiated
- 2013-07-01
- Posted
- 2013-08-07
- Terminated
- 2014-08-26
- Reason
- Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimension TACR Flex Reagent cartridge (DF107) lot BB4087 and its linked calibrator (DC 107) lot 3BD029. Falsely low Tacrolimus results across the assay range may lead to alterations in dosing.
- Root cause
- Process control
- Action
- Siemens sent an Urgent Medical Device Recall letter dated July 2013 to affected customers The letter identified the affected product, problem, potential risk, and actions to be taken by customers. Customers were requested to discuss the letter with thier laboratory director to determine if additional follow up is appropriate for patient tests conducted with the affected device lots. Additionally, customers were asked to discard any remaining inventory and to complete and return the field correction effectiveness check for replacement product. They were also asked to forward the letter to anyone wha had received this affected product. For questions contact your Siemens Technical Solutions Center or your local Siemens Technical support representative.
- Quantity
- 5423
- Distribution
- USA Nationwide Distribution including the states of: TX, NC, GA, TN, UT, NJ, MI, VA, AR, LA, MO, FL, AZ, CA, DC, PA, NH, WI, SD, SC, NM, MD, and IN.
- Lot, serial or model codes
- TACR Flex Reagent cartridge, material #10444938, (DF107) lot BB4087 with expiration date 3/28/14 and its linked calibrator, material #10445012, (DC 107) lot 3BD029 with expiration date 03/01/14.
- 510(k) numbers
- ["K060502"]
- PMA numbers
- not on file
- Product code
- MLM
- Firm FEI
- 2517506
- Firm city
- Newark
- Firm state
- DE
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28