Skip to the content

FDA Medical Device Recalls (openFDA)

Remel Europe Ltd.: Oxoid Agglutinating Sera, Salmonella 9-0 R30957301

Recall number
Z-1923-2026
Recalling firm
Remel Europe Ltd.
Product
Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
Status
Open, Classified
Initiated
2026-03-25
Posted
2026-04-22
Terminated
not on file
Reason
Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.
Root cause
Under Investigation by firm
Action
Firm began notifying consignees on March 25, 2026 via letters titled "Urgent: Medical Device Recall." Laboratory professionals are advised to review results and destroy any remaining inventory of affected lots. Affected lots should not be used for any clinical laboratory testing. Customers should complete and return the included acknowledgement form for replacement/credit to be processed. Notice should be passed on to all who need to be aware. Firm is investigating the root cause and will implement corrective and preventive actions to prevent future recurrence.
Quantity
117 US; 950 OUS
Distribution
Worldwide - US Nationwide distribution to the states of AL, CA, FL, IA, IL, IN, MI, MN, MO, NC, NE, OH, SD, TX, WA.
Lot, serial or model codes
UDI-DI 05056080500270 lots 6172600, 6172741, 6171393, 6150168, 6107767, 3766399, 3766400, 3780497, 3719592, 3709358, 6266130, 6253410, 6260215, 6222350, 6233859, 6233858
510(k) numbers
not on file
PMA numbers
not on file
Product code
GRM
Firm FEI
3003750284
Firm city
West Ken Dartford
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register