Merge Healthcare, Inc.: Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.
- Recall number
- Z-1923-2017
- Recalling firm
- Merge Healthcare, Inc.
- Product
- Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.
- Status
- Terminated
- Initiated
- 2015-03-30
- Posted
- 2017-05-02
- Terminated
- 2017-05-05
- Reason
- The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.
- Root cause
- Unknown/Undetermined by firm
- Action
- The recall was initiated by telephone on March 30, 2015. Follow-up phone calls were made July 6, 2015. The firm was unable to provide a telephone script. The recalling firm also issued letters dated March 28, 2016, via email on April 6, 2016, dated July 11, 2016. The letters contained the same content, notified the customer of the issue, and informed them a fix was available. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they have the affected ISA CO2 sensors at their facility and are they interested in accepting the fix. If they decline the fix, they are asked to provide the reason why. For further questions, please call (877) 741-5369.
- Quantity
- 23 sensors (PhaseIn Units)
- Distribution
- US Distribution was made to medical facilities located in AL, FL, ID, LA, MO, NC, OK, SC, TX, and UT. There was government distribution and no foreign/military distribution.
- Lot, serial or model codes
- ISA CO2 Sensor serial numbers: 107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435, 107476, 107509, 107510, 107556, 107659, 107700, 107707, 107768, 107776, 107831, 107832, 107836, 107839, and 107841.
- 510(k) numbers
- ["K082421"]
- PMA numbers
- not on file
- Product code
- DQK
- Firm FEI
- 1000116130
- Firm city
- Hartland
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28