Skip to the content

FDA Medical Device Recalls (openFDA)

Apatech Limited: Actifuse ABX/Actifuse MIS System A bone void filler intended only for orthopedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure.

Recall number
Z-1922-2016
Recalling firm
Apatech Limited
Product
Actifuse ABX/Actifuse MIS System A bone void filler intended only for orthopedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure.
Status
Terminated
Initiated
2015-08-18
Posted
not on file
Terminated
2018-05-31
Reason
Baxter Healthcare is issuing a voluntary recall for all lots of Actifuse ABX and Actifuse MIS System products with expiry before July 29, 2017 due to the possibility that the products may have endotoxin levels above specification criteria.
Root cause
Process control
Action
The firm, Baxter, sent an "Urgent Product Recall Notification" letter dated 8/18/2015 and a follow-up letter date 4/29/2016 to consignees via FedEx, overnight mail. Baxter also notified affected customers by phone. The letters described the product, problem and actions to be take. Baxter Healthcare has instructed customers to do the following: 1. Locate and remove all affected product from their facilities. 2. Contact Baxter Healthcare Center for Service to arrange for return and credit at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday. 3. Complete the attached customer reply form and return it to Baxter by either fax it to 224-270-5457 or scanned e-mail to fca@baxter.com. 4. If you distribute this product to other facilities or departments with your institution, notify all affected parties of this communication. For general questions regarding this communication, please call Baxter's Therapeutic Medical Affairs at 805-657-0007 or 862-432-7941.
Quantity
11696 units
Distribution
Worldwide Distribution: United States: Nationwide (including Puerto Rico) and countries of: Canada, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France, Great Britain, Ireland, Italy, The Netherlands, Norway, Sweden, Slovenia, Czech Republic, Kuwait, Mexico, Brazil, Australia, Taiwan, and New Zealand.
Lot, serial or model codes
Product Code: 506005078050 Lot #: All Expiration Date: August 01, 2015 - July 29, 2017  Product Code:506005078051 Lot #: All Expiration Date: August 01, 2015 - July 29, 2017  Product Code:506005078052 Lot #: All Expiration Date: August 01, 2015 - July 29, 2017  Product Code:506005078058 Lot #: All Expiration Date: August 01, 2015 - July 29, 2017  Product Code:506005078060 Lot #: All Expiration Date: August 01, 2015 - July 29, 2017  Product Code:506005078070 Lot #: All Expiration Date: August 01, 2015 - July 29, 2017  Product Code:506005078072 Lot #: All Expiration Date: August 01, 2015 - July 29, 2017
510(k) numbers
["K071206","K081979","K082575","K090850"]
PMA numbers
not on file
Product code
MQV
Firm FEI
3004450973
Firm city
Elstree
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register