FDA Medical Device Recalls (openFDA)
Nestle HealthCare Nutrition: Nestle Nutrition, Compat, Enteral Feeding Tube, sterile, 8 French (REF 08310400), 10 French (REF 08310600), and 12 French (REF 08310500). Product Usage: The Compat Nasogastric Feeding Tube is indicated for the administration of nutrition, fluids and medications into the stomach r small bowel by the naso-enteric, route for patients who have an intact gastrointestinal tract but are physically unable to ingest enough nutrients orally to meet their current needs.
- Recall number
- Z-1922-2013
- Recalling firm
- Nestle HealthCare Nutrition
- Product
- Nestle Nutrition, Compat, Enteral Feeding Tube, sterile, 8 French (REF 08310400), 10 French (REF 08310600), and 12 French (REF 08310500). Product Usage: The Compat Nasogastric Feeding Tube is indicated for the administration of nutrition, fluids and medications into the stomach r small bowel by the naso-enteric, route for patients who have an intact gastrointestinal tract but are physically unable to ingest enough nutrients orally to meet their current needs.
- Status
- Terminated
- Initiated
- 2013-07-24
- Posted
- 2013-08-06
- Terminated
- 2014-06-13
- Reason
- Nestle HealthCare Nutrition, Inc. is initiating a recall on some of Compat NG Tubes because they have been delivered in an insufficiently sealed pouch, thus compromising the sterility claimed on the label.
- Root cause
- Packaging process control
- Action
- Nestle Health Science sent a "Medical Device Recall" letter dated July 24, 2013. The letter was addressed to Risk Manger. The letter described the product involved, problem and actions to be taken. Customers were advised that products currently in use should not be returned. Customers were requested to examine their inventory, quarantine the product, carry the recall to the user facility level and to complete and return the Acknowledgment / Response Form. For questions call Nestle HealthCare Nutrition Customer Service at 1-877-463-7853.
- Quantity
- 37,310
- Distribution
- US Nationwide Distribution - including the states of: AK, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA.
- Lot, serial or model codes
- REF 08310400 - Compat¿ Nasogastric Feeding Tube 8 French (8Fr) in the following lots: 01413G 02412K 08211Z 08712W 14212U 14611P 21512Y 35411H REF 08310500- Compat¿ Nasogastric Feeding Tube 12 French (12Fr) in the following lots: 01211AV 01513N 04111D, 04111P, 04111V 05812F 05912F 06313N 21512G 23412BV 34712V 35411U REF 08310600 - Compat¿ Nasogastric Feeding Tube 10 French (10Fr) in the following lots: 08111M 08712D 10211E 13712W 21512F 05312N
- 510(k) numbers
- ["K915439"]
- PMA numbers
- not on file
- Product code
- FPD
- Firm FEI
- 2110851
- Firm city
- Minnetonka
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28