FDA Medical Device Recalls (openFDA)
Roho Inc.: ROHO HIGH PROFILE Single Compartment Cushion with Sensor Ready Technology and ROHO MID PROFILE Single Compartment Cushion with Sensor Ready Technology.
- Recall number
- Z-1921-2017
- Recalling firm
- Roho Inc.
- Product
- ROHO HIGH PROFILE Single Compartment Cushion with Sensor Ready Technology and ROHO MID PROFILE Single Compartment Cushion with Sensor Ready Technology.
- Status
- Terminated
- Initiated
- 2017-03-20
- Posted
- 2017-04-05
- Terminated
- 2019-05-10
- Reason
- When the Smart Check tool is used to perform a Smart Setup with one of the affected cushions, it may adversely affect the setup process resulting in an improper cushion setup. Sitting on an over inflated or under inflated Sensor Ready Cushion may reduce or eliminate the benefits provided by the cushion, increasing the risk to skin or to other soft tissue.
- Root cause
- Equipment maintenance
- Action
- On 3/13/17, firm began notifying customers of recall via emailed or mailed letter. The notification sent to dealers and distributors includes a request to forward the communication to their customer and send an acknowledgement back once all of their customers have received a copy of the communication. Customers (product users) will not be required to send the product they are using back. Instead, the recall letter provided two options: 1) keep the cushion and follow additional set-up instructions provided within the letter that would mitigate the risk or 2) contact the supplier to have the cushion replaced. Product that had been in use and is replaced will be destroyed (evidence to be provided to ROHO, Inc. if destroyed by a customer). Unused product that is still in stock with the firm will be destroyed.
- Quantity
- 5646 units
- Distribution
- Nationwide, Sweden, Germany, New Zealand, Finland, Australia, Singapore, Korea, Luxembourg, Czech Republic, South Africa, Canada, Switzerland, United Arab Emirates, Spain, Austria, Hong Kong, Iceland, Japan, Canada, Italy, Ireland, France, United Kingdom
- Lot, serial or model codes
- Serial Numbers: A2920163 through A3185287 Model Numbers: 1R1010H-CA-SC, 1R109M-CA-SR, 1R1211H-CH-SR, 1R88H-CA-SR, 1R1010H-CA-SCDM, 1R109M-CF-SR, 1R1212H-CA-SR, 1R88H-CF-SR, 1R1010H-CA-SR, 1R109M-CH-SR, 1R1212H-CH-SR, 1R88H-CH-SR, 1R1010H-CF-SCDM, 1R1110H-CA-SC, 1R129H-CA-SR, 1R88M-CA-SR, 1R1010H-CF-SR, 1R1110H-CA-SR, 1R1310H-CA-SR, 1R88M-CF-SR, 1R1010H-CH-SR, 1R1110H-CF-SR, 1R1310H-CF-SR, 1R89H-CA-SR, 1R1010M-CA-SC, 1R1110H-CH-SC, 1R1311H-CA-SR, 1R89H-CF-SR, 1R1010M-CA-SCDM, 1R1110H-CH-SR, 1R1311H-CH-SR, 1R89M-CA-SR, 1R1010M-CA-SR, 1R1110M-CA-SR, 1R1411H-CA-SR, 1R89M-CF-SR, 1R1010M-CF-SCDM, 1R1110M-CF-SR, 1R1411H-CF-SR, 1R910H-CA-SR, 1R1010M-CF-SR, 1R1110M-CH-SC, 1R1511H-CA-SR, 1R910H-CF-SR, 1R1010M-CH-SR, 1R1110M-CH-SR, 1R66H-CA-SR, 1R910H-CH-SC, 1R1011H-CA-SC, 1R1111H-CA-SC, 1R66H-CF-SR, 1R910H-CH-SR, 1R1011H-CA-SR, 1R1111H-CA-SR, 1R67H-CF-SR, 1R910M-CA-SR, 1R1011H-CF-SR, 1R1111H-CF-SR, 1R77H-CA-SR, 1R910M-CF-SR, 1R1011H-CH-SC, 1R1111H-CH-SC, 1R77H-CF-SR, 1R911H-CA-SC, 1R1011H-CH-SR, 1R1111H-CH-SR, 1R77M-CA-SR, 1R911H-CA-SR, 1R1011M-CA-SC, 1R1111M-CA-SR, 1R77M-CF-SR, 1R911H-CF-SR, 1R1011M-CA-SR, 1R1111M-CF-SR, 1R78H-CA-SR, 1R911H-CH-SR, 1R1011M-CF-SR, 1R1111M-CH-SR, 1R78H-CF-SR, 1R98H-CA-SR, 1R1012H-CA-SR, 1R1112H-CA-SR, 1R78M-CA-SR, 1R98H-CF-SR, 1R1012H-CF-SR, 1R1112H-CF-SR, 1R78M-CF-SR, 1R99H-CA-SC, 1R1012H-CH-SR, 1R1112H-CH-SR, 1R79H-CA-SR, 1R99H-CA-SCDM, 1R108H-CA-SR, 1R119H-CA-SC, 1R79H-CF-SR, 1R99H-CA-SR, 1R108H-CF-SR, 1R119H-CA-SR, 1R810H-CA-SR, 1R99H-CF-SCDM, 1R109H-CA-SCDM, 1R119H-CF-SR, 1R810H-CF-SR, 1R99H-CF-SR, 1R109H-CA-SR, 1R1210H-CA-SC, 1R810M-CA-SR, 1R99H-CH-SR, 1R109H-CF-SCDM, 1R1210H-CA-SR, 1R87H-CA-SC, 1R99M-CA-SR, 1R109H-CF-SR, 1R1210H-CF-SR, 1R87H-CA-SR, 1R99M-CF-SR, 1R109H-CH-SC, 1R1210H-CH-SR, 1R87H-CF-SR, 1R99M-CH-SR, 1R109H-CH-SR, 1R1211H-CA-SR
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KIC
- Firm FEI
- 3008630266
- Firm city
- Belleville
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28