ICU Medical, Inc.: REF: Z3806, 131" (333 cm) 10 Drop Primary Set w/2 MicroClave, Remv 4 Gang 4-way Stopcocks (1 Remv), Check Valve, Rotating Luer, 1 Ext, Sterile R, UDI: (01)00887709059108

Recall number
Z-1920-2020
Recalling firm
ICU Medical, Inc.
Product
REF: Z3806, 131" (333 cm) 10 Drop Primary Set w/2 MicroClave, Remv 4 Gang 4-way Stopcocks (1 Remv), Check Valve, Rotating Luer, 1 Ext, Sterile R, UDI: (01)00887709059108
Status
Terminated
Initiated
2019-10-04
Posted
not on file
Terminated
2021-12-28
Reason
There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.
Root cause
Mixed-up of materials/components
Action
On October 4, 2019, a "Urgent: Medical Device Recall" Letter was sent to all affected customers via US Mail. In addition to providing customers information on the recalled products, the letter asked customers to do the following: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) Return affected product using the return label provided with this letter. Contact Stericycle at 1-866-367-4703 (Monday - Friday, 8:00 am-5:00 pm ET) if you have not received a return label or require additional labels for returning the affected product. The return labels are for single use only. Please do not reproduce. Please visit http://expertezlabel.com to request additional labels for returning affected product. To ensure proper and timely credit, follow the instructions on the return label for returning product. Upon receipt of the completed response form and return of the affected product, ICU Medical will credit you for any product returned. You will only receive credit for product that you return. NOTE: Credits for product purchased through distributor will be credited by the distributor. 4) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product / Table 1 sections of this notification and ask them to contact Stericycle at 1-866-367-4703 (M-F, 8am-5pm ET) to obtain a response form. 5) Replacement product is available. Please contact customer service using the information provided below. - Global Complaint Management - 1-844-654-7780 or ProductComplaintsPP@icumed.com, To report adverse events or product complaints - C
Quantity
50 units
Distribution
US:AK,AL,AR,AZ,CA,CO,CT,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,SD,TN,TX,UT,VA,WA,WI,WY OUS: Canada *** "REF" numbers AH7117 and B30211 were not marketed or distributed within the US***
Lot, serial or model codes
Lot number: 4169642
510(k) numbers
["K964435"]
PMA numbers
not on file
Product code
FMG
Firm FEI
2025816
Firm city
San Clemente
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register