FDA Medical Device Recalls (openFDA)
Medtronic Neurosurgery: Medtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Catalog No. 46913. Intended to allow drainage and measurement of cerebrospinal fluid (CSF) and other fluids of similar physical character from the ventricles or the lumbar subarachnoid (intrathecal) region.
- Recall number
- Z-1920-2014
- Recalling firm
- Medtronic Neurosurgery
- Product
- Medtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Catalog No. 46913. Intended to allow drainage and measurement of cerebrospinal fluid (CSF) and other fluids of similar physical character from the ventricles or the lumbar subarachnoid (intrathecal) region.
- Status
- Terminated
- Initiated
- 2014-06-09
- Posted
- 2014-06-27
- Terminated
- 2015-01-14
- Reason
- Medtronic Neurosurgery is recalling the Medtronic Duet External Drainage and Monitoring Systems because the patient line tubing in the Duet External Drainage and Monitoring Systems may become disconnected from the patient line stopcock.
- Root cause
- Device Design
- Action
- An Urgent Medical Device Recall letter, dated 6/9/14, was sent to all their customers who purchased the Medtronic Duet External Drainage and Monitoring Systems. The letter informs the customers of the problem identified and the actions to be taken. Customers are instructed to cease use of the affected product, return any used units, account for any product used and verify that connections are in the condition as stated in the Instructions for Use for product currently in use. Additionally, customers are to complete the Customer Product Accountability Form. Customers with questions about the recall letter are instructed to contact Bob Shokoohi at (805) 571-8725.
- Quantity
- 3,026 units
- Distribution
- Worldwide Distribution -- USA, Australia, Belgium, Canada, Croatia, Cyprus, Denmark, Egypt, Finland, France, Greece, Iran, Israel, Italy, Kuwait, Netherlands, Norway, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Switzerland, Turkey, and United Kingdom.
- Lot, serial or model codes
- Lot No. 206923217 207096627 207096630 207187691 207224897 207269986 207466017 207565424 207565425 207659575 207659576 207739874 207875274 207945037 207982847
- 510(k) numbers
- ["K984053"]
- PMA numbers
- not on file
- Product code
- JXG
- Firm FEI
- 3015531529
- Firm city
- Goleta
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28