Agilent Technologies: Brand Name: Dako CoverStainer, Dako CoverStainer CPO Product Name: CoverStainer, CoverStainer CPO Model/Catalog Number: CS10030, CS10040, CS10026, CS10026E Software Version: N/A Product Description: Microscope Slide Stainer IVD Component: N/A
- Recall number
- Z-1919-2024
- Recalling firm
- Agilent Technologies
- Product
- Brand Name: Dako CoverStainer, Dako CoverStainer CPO Product Name: CoverStainer, CoverStainer CPO Model/Catalog Number: CS10030, CS10040, CS10026, CS10026E Software Version: N/A Product Description: Microscope Slide Stainer IVD Component: N/A
- Status
- Open, Classified
- Initiated
- 2024-04-02
- Posted
- 2024-05-29
- Terminated
- not on file
- Reason
- There is the potential for the front plexiglass door of the automated slide processing system to break causing potential injury.
- Root cause
- Device Design
- Action
- On April 15, 2024, Agilent Technologies, issued a "Urgent Medical Device Correction" notification to affected consignees. Agilent asked consignees to take the following actions: 1. Monitor your device and contact the service organization in case you discover a problem with the plexiglass on the lower front door. - In case the plexiglass of the lower front door is broken, immediately cease operation of the instrument - Do not remove the lower front door assembly, and do not run the instrument without the cover2.Confirm that you have received this information by signing and returning enclosed acknowledgement form Agilent Agilent is currently improving the design of the assembly to correct the issue. When the new design becomes available, the Agilent Service Organization will contact you and update your device at the next service visit. Please document the information on the enclosed Acknowledgement Form and send to Agilent s Field Action Team at fieldactions@agilent.com. In case of any question, please contact your local sales representative or Agilent service organization. We kindly ask you to inform those who need to be aware of this notification within your organization or any other organization to which the affected product(s) have been transferred. Please ensure that your organization maintains awareness of this notice. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
- Quantity
- 1160 devices
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, TN, MI, MD, MA, OR, AK, MO, NC, NH, GA, RI and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, Hong Kong, Spain, Italy, Japan, Malaysia, Netherland, New Zealand, Norway, Peru, South Korea, Sweden, Switzerland, Taiwan, The United Kingdom, Algeria, Bulgaria, Costa Rica, Cyrpus, Czech Republic, Ecuador, Egypt, Estonia, Greece, Hungary, India, Iraq, Ireland, Isreal, Kenya, Kuwait, Latvia, Lebanon, Morocco, Oman, Poland, Portugal, Qatar, Russia, Saudi Arabia, Slovakia, Thailand, Turkey, Ukraine, United Arab Emirates, Vietnam.
- Lot, serial or model codes
- Lot Code: All Lots
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KPA
- Firm FEI
- 1718499
- Firm city
- Loveland
- Firm state
- CO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28