Skip to the content

FDA Medical Device Recalls (openFDA)

LivaNova USA: Inspire 6 Integrated Phisio OXY w/ HVR Reservoir, Sterile EO, Catalog # 050713 The intended uses for the two elements that constitute the oxygenator/reservoir integrated devices are: INSPIRE 6M Hollow Fiber Oxygenator: The INSPIRE 6M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. The INSPIRE 6M is intended to be used for 6 hours or less. INSPIRE HVR (Hardshell Venous/Cardiotomy Reservoir): The INSPIRE HVR is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, de-foams and filters venous blood and suction blood. INSPIRE HVR can be used post-operatively for chest drainage. The INSPIRE HVR is intended to be used for 6 hours or less

Recall number
Z-1918-2018
Recalling firm
LivaNova USA
Product
Inspire 6 Integrated Phisio OXY w/ HVR Reservoir, Sterile EO, Catalog # 050713 The intended uses for the two elements that constitute the oxygenator/reservoir integrated devices are: INSPIRE 6M Hollow Fiber Oxygenator: The INSPIRE 6M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. The INSPIRE 6M is intended to be used for 6 hours or less. INSPIRE HVR (Hardshell Venous/Cardiotomy Reservoir): The INSPIRE HVR is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, de-foams and filters venous blood and suction blood. INSPIRE HVR can be used post-operatively for chest drainage. The INSPIRE HVR is intended to be used for 6 hours or less
Status
Terminated
Initiated
2018-03-13
Posted
not on file
Terminated
not on file
Reason
Some integrated sterile INSPIRE family products contain an incorrect expiration date printed in the Unique Device Identification (UDI) bar code and readable code just beneath the barcode. Specifically, the expiration date is reported as DDMMYY instead of YYMMDD. The expiration date in the applicable field on the package labels are correct as printed.
Root cause
Incorrect or no expiration date
Action
On March 13, 2018, notification letters were emailed. Customers were instructed to return response forms, which indicated the following: amount of unused product, and whether they would contact the firm to relabel effected products or if they preferred to be trained on the relabeling procedure. If product was further distributed, customers were instructed to send the notification letter to these additional customers within 2 business days. No action is required for product already used since it was consumed prior to the true expiration date. For further questions, please call (281) 228-7260.
Quantity
42
Distribution
US Distribution to the states of : NJ, CA, MN, MO, MA
Lot, serial or model codes
Lot: 1709190228, UDI: (01)08033178112352(17)200917(10)1709190228, Exp: 09/17/2020
510(k) numbers
["K130209"]
PMA numbers
not on file
Product code
DTZ
Firm FEI
1718850
Firm city
Arvada
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register