Celsus Laboratories Inc: Celsus Laboratories Heparin Lithium Lyophilized; Lithium Heparin for use in blood collection tube devices and in vitro diagnostic devices.
- Recall number
- Z-1917-2008
- Recalling firm
- Celsus Laboratories Inc
- Product
- Celsus Laboratories Heparin Lithium Lyophilized; Lithium Heparin for use in blood collection tube devices and in vitro diagnostic devices.
- Status
- Terminated
- Initiated
- 2008-03-31
- Posted
- 2008-07-18
- Terminated
- 2012-09-21
- Reason
- Contamination: The firm learned that up to 2.2% oversulfated chondroitin was found in a batch of crude heparin they received from their supplier.
- Root cause
- Nonconforming Material/Component
- Action
- Celsus notified their customers by e-mail on April 24, 2008, explaining that oversulfated chondroitin was found in a batch of crude heparin they received from their supplier. Celsus recommended the device manufacturers quarantine material from the specified lots.
- Quantity
- HL-31707=11,542g, HL-31807=12,836g, HL-32708=8,374g, HL-32808=12,128g & HL-32908=5,000g
- Distribution
- Worldwide Distribution. The Lithium Heparin was sold to companies in USA (Nebraska, New Jersey, Ohio), Puerto Rico, Austria, and Denmark.
- Lot, serial or model codes
- Lot# HL-31707, HL-31807, HL-32708, HL-32808 & HL-32908
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JKA
- Firm FEI
- 1527761
- Firm city
- Cincinnati
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28