Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes (USA) Products LLC: STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010

Recall number
Z-1916-2024
Recalling firm
Synthes (USA) Products LLC
Product
STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010
Status
Open, Classified
Initiated
2024-04-25
Posted
2024-05-28
Terminated
not on file
Reason
Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay
Root cause
Process change control
Action
DePuy Synthes issued Urgent Medical Device Recall (Removal) Letter on 4/26/24. Letter states reason for recall, health risk and action to take: Examine your inventory to determine if you have the subject product lots and quarantine them immediately. DO NOT USE THE SUBJECT PRODUCTS. 2. Contact your DePuy Synthes Sales Consultant or contact the customer support services at ra-dpyuscsspecial@ its.jnj.com to coordinate the return/credits of the subject product lots. 3. Review, complete, sign, and return the attached Business Response Form (page 3 of this letter) to OneMDField- Actions@its.jnj.com within three (3) business days of receipt of this notification. Please include in the email subject: FA 2363331. 4. Please complete the attached Business Response Form even if you do not have the subject product lots on hand. 5. Forward this notice to anyone in your facility that needs to be informed (e.g., those who manage, transport, store, stock, or use the subject products). 6. If any of the subject product lots have been forwarded to another facility, contact that facility and provide them with this notice. 7. Post a copy of this notice in a visible area for awareness and keep a copy for your records. If you have any questions, please contact your local DePuy Synthes Sales Consultant. For Medical Information request, please visit our website: https://www.jnjmedicaldevices.com/mir.
Quantity
261 units
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Lot, serial or model codes
UDI-DI: 10886982075192 Lot Numbers: 97P6250, 97P6251, 97P6252
510(k) numbers
not on file
PMA numbers
not on file
Product code
HXX
Firm FEI
3008812560
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source